A depth profile fitting model for a commercial total reflection X-ray fluorescence spectrometer

Yoshihiro Mori, Kenichi Uemura, Kengo Shimanoe

Research output: Contribution to journalArticlepeer-review

6 Citations (Scopus)

Abstract

We have proposed a practical depth profiling model for a commercial Total Reflection X-Ray Fluorescence (TXRF) spectrometry instrument. The model includes three factors peculiar to a commercial TXRF instrument: (a) the irradiated X-ray photon density on the sample surface depends on a glancing angle, (b) the X-ray irradiated area becomes smaller than the detector view over a certain glancing angle, and (c) the incident X-ray has angular divergence. This model was optimized by comparing the measured Si-Ka for a silicon wafer with the calculated one. The results indicated that all of the three factors are indispensable for our instrument. We have applied this model to a surface contaminant and discussed its adequacy.

Original languageEnglish
Pages (from-to)823-828
Number of pages6
JournalSpectrochimica Acta - Part B Atomic Spectroscopy
Volume52
Issue number7
DOIs
Publication statusPublished - Jul 1 1997
Externally publishedYes

All Science Journal Classification (ASJC) codes

  • Analytical Chemistry
  • Atomic and Molecular Physics, and Optics
  • Instrumentation
  • Spectroscopy

Fingerprint Dive into the research topics of 'A depth profile fitting model for a commercial total reflection X-ray fluorescence spectrometer'. Together they form a unique fingerprint.