A Gastric Tube Monitoring System for Clinical Applications

Teruyoshi Sasayama, Yuji Gotoh, Keiji Enpuku

Research output: Contribution to journalArticle

2 Citations (Scopus)

Abstract

To correctly insert a gastric tube into the stomach, the position of an attached permanent magnet is estimated by solving the inverse problem using measurement data obtained from magneto-impedance (MI) sensors placed outside the patient's body. The ubiquity and power of tablet computers make them promising tools to easily confirm the position of the tip of the gastric tube. However, a single mobile device may be insufficient for actually solving the inverse problem. Therefore, we propose a distributed cooperative local access network system for the estimation of the position of a permanent magnet attached to the tip of a gastric tube. In addition, we evaluate whether ferromagnetic materials near a sensor affect the performance of the system. Experimental results show that the position estimation error is less than 10 mm, even when the tip is 150 mm from a MI sensor, and that the tip position can be displayed on a tablet computer every 0.25 s. These results suggest that the proposed method holds promise for hospital applications.

Original languageEnglish
Article number7946121
JournalIEEE Transactions on Magnetics
Volume53
Issue number11
DOIs
Publication statusPublished - Nov 1 2017

Fingerprint

Inverse problems
Permanent magnets
Tablets
Monitoring
Sensors
Ferromagnetic materials
Mobile devices
Error analysis

All Science Journal Classification (ASJC) codes

  • Electronic, Optical and Magnetic Materials
  • Electrical and Electronic Engineering

Cite this

A Gastric Tube Monitoring System for Clinical Applications. / Sasayama, Teruyoshi; Gotoh, Yuji; Enpuku, Keiji.

In: IEEE Transactions on Magnetics, Vol. 53, No. 11, 7946121, 01.11.2017.

Research output: Contribution to journalArticle

@article{b762100f4940491d9b9f2c9cc1bbd768,
title = "A Gastric Tube Monitoring System for Clinical Applications",
abstract = "To correctly insert a gastric tube into the stomach, the position of an attached permanent magnet is estimated by solving the inverse problem using measurement data obtained from magneto-impedance (MI) sensors placed outside the patient's body. The ubiquity and power of tablet computers make them promising tools to easily confirm the position of the tip of the gastric tube. However, a single mobile device may be insufficient for actually solving the inverse problem. Therefore, we propose a distributed cooperative local access network system for the estimation of the position of a permanent magnet attached to the tip of a gastric tube. In addition, we evaluate whether ferromagnetic materials near a sensor affect the performance of the system. Experimental results show that the position estimation error is less than 10 mm, even when the tip is 150 mm from a MI sensor, and that the tip position can be displayed on a tablet computer every 0.25 s. These results suggest that the proposed method holds promise for hospital applications.",
author = "Teruyoshi Sasayama and Yuji Gotoh and Keiji Enpuku",
year = "2017",
month = "11",
day = "1",
doi = "10.1109/TMAG.2017.2714683",
language = "English",
volume = "53",
journal = "IEEE Transactions on Magnetics",
issn = "0018-9464",
publisher = "Institute of Electrical and Electronics Engineers Inc.",
number = "11",

}

TY - JOUR

T1 - A Gastric Tube Monitoring System for Clinical Applications

AU - Sasayama, Teruyoshi

AU - Gotoh, Yuji

AU - Enpuku, Keiji

PY - 2017/11/1

Y1 - 2017/11/1

N2 - To correctly insert a gastric tube into the stomach, the position of an attached permanent magnet is estimated by solving the inverse problem using measurement data obtained from magneto-impedance (MI) sensors placed outside the patient's body. The ubiquity and power of tablet computers make them promising tools to easily confirm the position of the tip of the gastric tube. However, a single mobile device may be insufficient for actually solving the inverse problem. Therefore, we propose a distributed cooperative local access network system for the estimation of the position of a permanent magnet attached to the tip of a gastric tube. In addition, we evaluate whether ferromagnetic materials near a sensor affect the performance of the system. Experimental results show that the position estimation error is less than 10 mm, even when the tip is 150 mm from a MI sensor, and that the tip position can be displayed on a tablet computer every 0.25 s. These results suggest that the proposed method holds promise for hospital applications.

AB - To correctly insert a gastric tube into the stomach, the position of an attached permanent magnet is estimated by solving the inverse problem using measurement data obtained from magneto-impedance (MI) sensors placed outside the patient's body. The ubiquity and power of tablet computers make them promising tools to easily confirm the position of the tip of the gastric tube. However, a single mobile device may be insufficient for actually solving the inverse problem. Therefore, we propose a distributed cooperative local access network system for the estimation of the position of a permanent magnet attached to the tip of a gastric tube. In addition, we evaluate whether ferromagnetic materials near a sensor affect the performance of the system. Experimental results show that the position estimation error is less than 10 mm, even when the tip is 150 mm from a MI sensor, and that the tip position can be displayed on a tablet computer every 0.25 s. These results suggest that the proposed method holds promise for hospital applications.

UR - http://www.scopus.com/inward/record.url?scp=85021823763&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85021823763&partnerID=8YFLogxK

U2 - 10.1109/TMAG.2017.2714683

DO - 10.1109/TMAG.2017.2714683

M3 - Article

VL - 53

JO - IEEE Transactions on Magnetics

JF - IEEE Transactions on Magnetics

SN - 0018-9464

IS - 11

M1 - 7946121