Active vibration suppression on an image of a scanning electron microscope

Koichi Matsuda, Natsuki Kawamura, Yoichi Kanemitsu, Shinya Kijimoto, Katsuhide Watanabe, Eiichi Izumi

Research output: Contribution to journalConference article

2 Citations (Scopus)

Abstract

This paper proposes a new approach to reducing an effect of floor vibration on an image of a scanning electron microscope. An image-shifting coil is used to move the electron probe in order to cancel undesirable motion of a specimen due to the floor vibration. The floor vibration is structurally transmitted through the microscope and detected by two acceleration sensors at the root of the specimen chamber of the microscope. The outputs of the acceleration sensors are fed forward into a controller to move the electron probe by the image-shifting coil. The feed-forward controllers are designed in two ways. The first one is based on a transfer function from the sensor outputs to the relative displacement of a specimen to the electron probe being at rest. The microscope is put on a table attached to a shaker. Sinusoidal excitation tests are done many times to estimate the transfer functions from vibrating images of a micro scale. Moreover, the second controller is designed by manually amplifying and delaying the sensor outputs so as to minimize amplitude of the vibrating images on a CRT. Those two controllers are implemented as a digital filter running on a digital signal processor.

Original languageEnglish
Pages (from-to)622-630
Number of pages9
JournalProceedings of SPIE - The International Society for Optical Engineering
Volume3998
Publication statusPublished - Jan 1 2000
EventMetrology, Inspection, and Process Control for Microlithography XIV - Santa Clara, CA, USA
Duration: Feb 28 2000Mar 2 2000

Fingerprint

vibration
Electron microscopes
electron microscopes
retarding
Scanning
controllers
electron probes
electron
Controllers
Microscopes
scanning
Sensors
sensor
sensors
microscopes
Transfer functions
probe
Electrons
transfer function
transfer functions

All Science Journal Classification (ASJC) codes

  • Electronic, Optical and Magnetic Materials
  • Condensed Matter Physics
  • Computer Science Applications
  • Applied Mathematics
  • Electrical and Electronic Engineering

Cite this

Active vibration suppression on an image of a scanning electron microscope. / Matsuda, Koichi; Kawamura, Natsuki; Kanemitsu, Yoichi; Kijimoto, Shinya; Watanabe, Katsuhide; Izumi, Eiichi.

In: Proceedings of SPIE - The International Society for Optical Engineering, Vol. 3998, 01.01.2000, p. 622-630.

Research output: Contribution to journalConference article

Matsuda, Koichi ; Kawamura, Natsuki ; Kanemitsu, Yoichi ; Kijimoto, Shinya ; Watanabe, Katsuhide ; Izumi, Eiichi. / Active vibration suppression on an image of a scanning electron microscope. In: Proceedings of SPIE - The International Society for Optical Engineering. 2000 ; Vol. 3998. pp. 622-630.
@article{fe21a5868f7b4d52b365c66f3f7e3201,
title = "Active vibration suppression on an image of a scanning electron microscope",
abstract = "This paper proposes a new approach to reducing an effect of floor vibration on an image of a scanning electron microscope. An image-shifting coil is used to move the electron probe in order to cancel undesirable motion of a specimen due to the floor vibration. The floor vibration is structurally transmitted through the microscope and detected by two acceleration sensors at the root of the specimen chamber of the microscope. The outputs of the acceleration sensors are fed forward into a controller to move the electron probe by the image-shifting coil. The feed-forward controllers are designed in two ways. The first one is based on a transfer function from the sensor outputs to the relative displacement of a specimen to the electron probe being at rest. The microscope is put on a table attached to a shaker. Sinusoidal excitation tests are done many times to estimate the transfer functions from vibrating images of a micro scale. Moreover, the second controller is designed by manually amplifying and delaying the sensor outputs so as to minimize amplitude of the vibrating images on a CRT. Those two controllers are implemented as a digital filter running on a digital signal processor.",
author = "Koichi Matsuda and Natsuki Kawamura and Yoichi Kanemitsu and Shinya Kijimoto and Katsuhide Watanabe and Eiichi Izumi",
year = "2000",
month = "1",
day = "1",
language = "English",
volume = "3998",
pages = "622--630",
journal = "Proceedings of SPIE - The International Society for Optical Engineering",
issn = "0277-786X",
publisher = "SPIE",

}

TY - JOUR

T1 - Active vibration suppression on an image of a scanning electron microscope

AU - Matsuda, Koichi

AU - Kawamura, Natsuki

AU - Kanemitsu, Yoichi

AU - Kijimoto, Shinya

AU - Watanabe, Katsuhide

AU - Izumi, Eiichi

PY - 2000/1/1

Y1 - 2000/1/1

N2 - This paper proposes a new approach to reducing an effect of floor vibration on an image of a scanning electron microscope. An image-shifting coil is used to move the electron probe in order to cancel undesirable motion of a specimen due to the floor vibration. The floor vibration is structurally transmitted through the microscope and detected by two acceleration sensors at the root of the specimen chamber of the microscope. The outputs of the acceleration sensors are fed forward into a controller to move the electron probe by the image-shifting coil. The feed-forward controllers are designed in two ways. The first one is based on a transfer function from the sensor outputs to the relative displacement of a specimen to the electron probe being at rest. The microscope is put on a table attached to a shaker. Sinusoidal excitation tests are done many times to estimate the transfer functions from vibrating images of a micro scale. Moreover, the second controller is designed by manually amplifying and delaying the sensor outputs so as to minimize amplitude of the vibrating images on a CRT. Those two controllers are implemented as a digital filter running on a digital signal processor.

AB - This paper proposes a new approach to reducing an effect of floor vibration on an image of a scanning electron microscope. An image-shifting coil is used to move the electron probe in order to cancel undesirable motion of a specimen due to the floor vibration. The floor vibration is structurally transmitted through the microscope and detected by two acceleration sensors at the root of the specimen chamber of the microscope. The outputs of the acceleration sensors are fed forward into a controller to move the electron probe by the image-shifting coil. The feed-forward controllers are designed in two ways. The first one is based on a transfer function from the sensor outputs to the relative displacement of a specimen to the electron probe being at rest. The microscope is put on a table attached to a shaker. Sinusoidal excitation tests are done many times to estimate the transfer functions from vibrating images of a micro scale. Moreover, the second controller is designed by manually amplifying and delaying the sensor outputs so as to minimize amplitude of the vibrating images on a CRT. Those two controllers are implemented as a digital filter running on a digital signal processor.

UR - http://www.scopus.com/inward/record.url?scp=0033707697&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0033707697&partnerID=8YFLogxK

M3 - Conference article

AN - SCOPUS:0033707697

VL - 3998

SP - 622

EP - 630

JO - Proceedings of SPIE - The International Society for Optical Engineering

JF - Proceedings of SPIE - The International Society for Optical Engineering

SN - 0277-786X