An application of CCD read-out technique to neutron distribution measurement using the self-activation method with a CsI scintillator plate

Akihiro Nohtomi, Ryosuke Kurihara, Hiroyuki Kinoshita, Soichiro Honda, Masaaki Tokunaga, Heita Uno, Kiyomitsu Shinsho, Genichiro Wakabayashi, Yusuke Koba, Junichi Fukunaga, Yoshiyuki Umezu, Yasuhiko Nakamura, Saiji Ohga

Research output: Contribution to journalArticle

3 Citations (Scopus)

Abstract

In our previous paper, the self-activation of an NaI scintillator had been successfully utilized for detecting photo-neutrons around a high-energy X-ray radiotherapy machine; individual optical pulses from the self-activated scintillator are read-out by photo sensors such as a photomultiplier tube (PMT). In the present work, preliminary observations have been performed in order to apply a direct CCD read-out technique to the self-activation method with a CsI scintillator plate using a Pu-Be source and a 10-MV linac. In conclusion, it has been revealed that the CCD read-out technique is applicable to neutron measurement around a high-energy X-ray radiotherapy machine with the self-activation of a CsI plate. Such application may provide a possibility of novel method for simple neutron dose-distribution measurement.

Original languageEnglish
Pages (from-to)21-23
Number of pages3
JournalNuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Volume832
DOIs
Publication statusPublished - Oct 1 2016

Fingerprint

neutron distribution
Charge coupled devices
Phosphors
scintillation counters
charge coupled devices
Neutrons
Chemical activation
Radiotherapy
activation
neutrons
radiation therapy
X rays
Photomultipliers
photomultiplier tubes
Laser pulses
x rays
dosage
energy
sensors
Sensors

All Science Journal Classification (ASJC) codes

  • Nuclear and High Energy Physics
  • Instrumentation

Cite this

An application of CCD read-out technique to neutron distribution measurement using the self-activation method with a CsI scintillator plate. / Nohtomi, Akihiro; Kurihara, Ryosuke; Kinoshita, Hiroyuki; Honda, Soichiro; Tokunaga, Masaaki; Uno, Heita; Shinsho, Kiyomitsu; Wakabayashi, Genichiro; Koba, Yusuke; Fukunaga, Junichi; Umezu, Yoshiyuki; Nakamura, Yasuhiko; Ohga, Saiji.

In: Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 832, 01.10.2016, p. 21-23.

Research output: Contribution to journalArticle

Nohtomi, Akihiro ; Kurihara, Ryosuke ; Kinoshita, Hiroyuki ; Honda, Soichiro ; Tokunaga, Masaaki ; Uno, Heita ; Shinsho, Kiyomitsu ; Wakabayashi, Genichiro ; Koba, Yusuke ; Fukunaga, Junichi ; Umezu, Yoshiyuki ; Nakamura, Yasuhiko ; Ohga, Saiji. / An application of CCD read-out technique to neutron distribution measurement using the self-activation method with a CsI scintillator plate. In: Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 2016 ; Vol. 832. pp. 21-23.
@article{dd3aeb36a16f4864b870484ef9451017,
title = "An application of CCD read-out technique to neutron distribution measurement using the self-activation method with a CsI scintillator plate",
abstract = "In our previous paper, the self-activation of an NaI scintillator had been successfully utilized for detecting photo-neutrons around a high-energy X-ray radiotherapy machine; individual optical pulses from the self-activated scintillator are read-out by photo sensors such as a photomultiplier tube (PMT). In the present work, preliminary observations have been performed in order to apply a direct CCD read-out technique to the self-activation method with a CsI scintillator plate using a Pu-Be source and a 10-MV linac. In conclusion, it has been revealed that the CCD read-out technique is applicable to neutron measurement around a high-energy X-ray radiotherapy machine with the self-activation of a CsI plate. Such application may provide a possibility of novel method for simple neutron dose-distribution measurement.",
author = "Akihiro Nohtomi and Ryosuke Kurihara and Hiroyuki Kinoshita and Soichiro Honda and Masaaki Tokunaga and Heita Uno and Kiyomitsu Shinsho and Genichiro Wakabayashi and Yusuke Koba and Junichi Fukunaga and Yoshiyuki Umezu and Yasuhiko Nakamura and Saiji Ohga",
year = "2016",
month = "10",
day = "1",
doi = "10.1016/j.nima.2016.06.078",
language = "English",
volume = "832",
pages = "21--23",
journal = "Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment",
issn = "0168-9002",
publisher = "Elsevier",

}

TY - JOUR

T1 - An application of CCD read-out technique to neutron distribution measurement using the self-activation method with a CsI scintillator plate

AU - Nohtomi, Akihiro

AU - Kurihara, Ryosuke

AU - Kinoshita, Hiroyuki

AU - Honda, Soichiro

AU - Tokunaga, Masaaki

AU - Uno, Heita

AU - Shinsho, Kiyomitsu

AU - Wakabayashi, Genichiro

AU - Koba, Yusuke

AU - Fukunaga, Junichi

AU - Umezu, Yoshiyuki

AU - Nakamura, Yasuhiko

AU - Ohga, Saiji

PY - 2016/10/1

Y1 - 2016/10/1

N2 - In our previous paper, the self-activation of an NaI scintillator had been successfully utilized for detecting photo-neutrons around a high-energy X-ray radiotherapy machine; individual optical pulses from the self-activated scintillator are read-out by photo sensors such as a photomultiplier tube (PMT). In the present work, preliminary observations have been performed in order to apply a direct CCD read-out technique to the self-activation method with a CsI scintillator plate using a Pu-Be source and a 10-MV linac. In conclusion, it has been revealed that the CCD read-out technique is applicable to neutron measurement around a high-energy X-ray radiotherapy machine with the self-activation of a CsI plate. Such application may provide a possibility of novel method for simple neutron dose-distribution measurement.

AB - In our previous paper, the self-activation of an NaI scintillator had been successfully utilized for detecting photo-neutrons around a high-energy X-ray radiotherapy machine; individual optical pulses from the self-activated scintillator are read-out by photo sensors such as a photomultiplier tube (PMT). In the present work, preliminary observations have been performed in order to apply a direct CCD read-out technique to the self-activation method with a CsI scintillator plate using a Pu-Be source and a 10-MV linac. In conclusion, it has been revealed that the CCD read-out technique is applicable to neutron measurement around a high-energy X-ray radiotherapy machine with the self-activation of a CsI plate. Such application may provide a possibility of novel method for simple neutron dose-distribution measurement.

UR - http://www.scopus.com/inward/record.url?scp=84975744765&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84975744765&partnerID=8YFLogxK

U2 - 10.1016/j.nima.2016.06.078

DO - 10.1016/j.nima.2016.06.078

M3 - Article

AN - SCOPUS:84975744765

VL - 832

SP - 21

EP - 23

JO - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment

JF - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment

SN - 0168-9002