Automated MTF measurement in CT images with a simple wire phantom

Choirul Anam, Toshioh Fujibuchi, Freddy Haryanto, Wahyu Setia Budi, Heri Sutanto, Kusworo Adi, Zaenul Muhlisin, Geoff Dougherty

Research output: Contribution to journalArticle

Abstract

This study developed a simple wire phantom and an algorithm to automatically measure the modulation transfer function (MTF) in computed tomography (CT) and implemented it to evaluate the effect of focal spot size and reconstruction filter type. The phantom consisted of a resin cylinder filled with water, with a tin wire of diameter 0.1 mm positioned along the center of the cylinder. The automated MTF algorithm used an axial image of the phantom and comprised several steps. The center position of a region of interest (ROI) was automatically determined at the center of the wire image. The pixels were then summed along the y-direction to obtain the profile of the pixel values at a point along the x-direction. Following this, both edges of the profile were made equal to zero. The profile curve was then normalized so that the total of all the data was equal to unity. The normalized profile curve is the line spread function (LSF), and the MTF curve was obtained by taking its Fourier transform. Our system (phantom and algorithm) is able to differentiate the MTFs of CT images from different focal sizes and reconstruction filter types.

Original languageEnglish
Pages (from-to)179-187
Number of pages9
JournalPolish Journal of Medical Physics and Engineering
Volume25
Issue number3
DOIs
Publication statusPublished - Sep 1 2019

Fingerprint

Tomography
Tin
Fourier Analysis
Water
Direction compound

All Science Journal Classification (ASJC) codes

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

Automated MTF measurement in CT images with a simple wire phantom. / Anam, Choirul; Fujibuchi, Toshioh; Haryanto, Freddy; Budi, Wahyu Setia; Sutanto, Heri; Adi, Kusworo; Muhlisin, Zaenul; Dougherty, Geoff.

In: Polish Journal of Medical Physics and Engineering, Vol. 25, No. 3, 01.09.2019, p. 179-187.

Research output: Contribution to journalArticle

Anam, C, Fujibuchi, T, Haryanto, F, Budi, WS, Sutanto, H, Adi, K, Muhlisin, Z & Dougherty, G 2019, 'Automated MTF measurement in CT images with a simple wire phantom', Polish Journal of Medical Physics and Engineering, vol. 25, no. 3, pp. 179-187. https://doi.org/10.2478/pjmpe-2019-0024
Anam, Choirul ; Fujibuchi, Toshioh ; Haryanto, Freddy ; Budi, Wahyu Setia ; Sutanto, Heri ; Adi, Kusworo ; Muhlisin, Zaenul ; Dougherty, Geoff. / Automated MTF measurement in CT images with a simple wire phantom. In: Polish Journal of Medical Physics and Engineering. 2019 ; Vol. 25, No. 3. pp. 179-187.
@article{d9a37a0a35d0471580df270d65824130,
title = "Automated MTF measurement in CT images with a simple wire phantom",
abstract = "This study developed a simple wire phantom and an algorithm to automatically measure the modulation transfer function (MTF) in computed tomography (CT) and implemented it to evaluate the effect of focal spot size and reconstruction filter type. The phantom consisted of a resin cylinder filled with water, with a tin wire of diameter 0.1 mm positioned along the center of the cylinder. The automated MTF algorithm used an axial image of the phantom and comprised several steps. The center position of a region of interest (ROI) was automatically determined at the center of the wire image. The pixels were then summed along the y-direction to obtain the profile of the pixel values at a point along the x-direction. Following this, both edges of the profile were made equal to zero. The profile curve was then normalized so that the total of all the data was equal to unity. The normalized profile curve is the line spread function (LSF), and the MTF curve was obtained by taking its Fourier transform. Our system (phantom and algorithm) is able to differentiate the MTFs of CT images from different focal sizes and reconstruction filter types.",
author = "Choirul Anam and Toshioh Fujibuchi and Freddy Haryanto and Budi, {Wahyu Setia} and Heri Sutanto and Kusworo Adi and Zaenul Muhlisin and Geoff Dougherty",
year = "2019",
month = "9",
day = "1",
doi = "10.2478/pjmpe-2019-0024",
language = "English",
volume = "25",
pages = "179--187",
journal = "Polish Journal of Medical Physics and Engineering",
issn = "1425-4689",
publisher = "Polish Society of Medical Physics",
number = "3",

}

TY - JOUR

T1 - Automated MTF measurement in CT images with a simple wire phantom

AU - Anam, Choirul

AU - Fujibuchi, Toshioh

AU - Haryanto, Freddy

AU - Budi, Wahyu Setia

AU - Sutanto, Heri

AU - Adi, Kusworo

AU - Muhlisin, Zaenul

AU - Dougherty, Geoff

PY - 2019/9/1

Y1 - 2019/9/1

N2 - This study developed a simple wire phantom and an algorithm to automatically measure the modulation transfer function (MTF) in computed tomography (CT) and implemented it to evaluate the effect of focal spot size and reconstruction filter type. The phantom consisted of a resin cylinder filled with water, with a tin wire of diameter 0.1 mm positioned along the center of the cylinder. The automated MTF algorithm used an axial image of the phantom and comprised several steps. The center position of a region of interest (ROI) was automatically determined at the center of the wire image. The pixels were then summed along the y-direction to obtain the profile of the pixel values at a point along the x-direction. Following this, both edges of the profile were made equal to zero. The profile curve was then normalized so that the total of all the data was equal to unity. The normalized profile curve is the line spread function (LSF), and the MTF curve was obtained by taking its Fourier transform. Our system (phantom and algorithm) is able to differentiate the MTFs of CT images from different focal sizes and reconstruction filter types.

AB - This study developed a simple wire phantom and an algorithm to automatically measure the modulation transfer function (MTF) in computed tomography (CT) and implemented it to evaluate the effect of focal spot size and reconstruction filter type. The phantom consisted of a resin cylinder filled with water, with a tin wire of diameter 0.1 mm positioned along the center of the cylinder. The automated MTF algorithm used an axial image of the phantom and comprised several steps. The center position of a region of interest (ROI) was automatically determined at the center of the wire image. The pixels were then summed along the y-direction to obtain the profile of the pixel values at a point along the x-direction. Following this, both edges of the profile were made equal to zero. The profile curve was then normalized so that the total of all the data was equal to unity. The normalized profile curve is the line spread function (LSF), and the MTF curve was obtained by taking its Fourier transform. Our system (phantom and algorithm) is able to differentiate the MTFs of CT images from different focal sizes and reconstruction filter types.

UR - http://www.scopus.com/inward/record.url?scp=85072090859&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85072090859&partnerID=8YFLogxK

U2 - 10.2478/pjmpe-2019-0024

DO - 10.2478/pjmpe-2019-0024

M3 - Article

AN - SCOPUS:85072090859

VL - 25

SP - 179

EP - 187

JO - Polish Journal of Medical Physics and Engineering

JF - Polish Journal of Medical Physics and Engineering

SN - 1425-4689

IS - 3

ER -