Axial and lateral displacement measurements of a microsphere based on the critical-angle method

Eiji Higurashi, Renshi Sawada, Takahiro Ito

Research output: Contribution to journalArticlepeer-review

5 Citations (Scopus)

Abstract

A method has been developed for optically measuring nanometer-scale displacements of transparent and metal-coated microspheres in both the axial (vertical) and lateral (horizontal) directions. This method works by detecting changes in the internal reflection of a laser beam reflected from the microsphere after it passes through critical-angle prisms. For weakly reflective 10-μm-diameter polystyrene microspheres in water (relative refractive index n = 1.2), the detection resolutions as estimated from the full-width at half maximum (FWHM) noise level in the frequency region above 500 Hz in the axial and lateral directions were experimentally found to be as good as 1.7 and 1.1 nm, respectively. Furthermore, the lateral displacement resolution (0.2 nm) of a 10-μm-diameter metal-coated microsphere was better than the axial displacement resolution (0.9 nm) of a conventional flat mirror by more than a factor of four.

Original languageEnglish
Pages (from-to)4191-4196
Number of pages6
JournalJapanese Journal of Applied Physics, Part 1: Regular Papers and Short Notes and Review Papers
Volume37
Issue number7
DOIs
Publication statusPublished - Jul 1998
Externally publishedYes

All Science Journal Classification (ASJC) codes

  • Engineering(all)
  • Physics and Astronomy(all)

Fingerprint

Dive into the research topics of 'Axial and lateral displacement measurements of a microsphere based on the critical-angle method'. Together they form a unique fingerprint.

Cite this