C216 PVT MEASUREMENTS OF HIGH PRESSURE GAS BY THE BURNETT METHOD(Hydrogen and Reforming-2)

Naoya Sakoda, Kenta Shindo, Kan'ei Shinzato, Masamichi Kohno, Yasuyuki Takata, Motoo Fujii

Research output: Contribution to journalArticlepeer-review

Abstract

A PVT measurement apparatus by the Burnett method for the temperature range from room temperature to 250 ℃ and at pressures up to 100 MPa has been developed with a remote operation system. PVT measurements of hydrogen, helium and nitrogen have been accomplished up to 97 MPa. The density uncertainty of the present measurements at 80 ℃ is estimated to be 0.2 % from the deviations for nitrogen. An existing equation of state (EOS) for hydrogen agrees well with the present measurements within ±0.22 % at 80 ℃. The systematic deviations between the EOS and the present measurements at 120 ℃ and 160 ℃ are shown with the maximum deviation of 0.52 % at 120 ℃ and 97 MPa.
Original languageEnglish
Pages (from-to)_2-265_-_2-270_
JournalThe Proceedings of the International Conference on Power Engineering (ICOPE)
Volume2009
Issue number0
DOIs
Publication statusPublished - 2009

Fingerprint Dive into the research topics of 'C216 PVT MEASUREMENTS OF HIGH PRESSURE GAS BY THE BURNETT METHOD(Hydrogen and Reforming-2)'. Together they form a unique fingerprint.