TY - JOUR
T1 - Comparative study of two different chromatographic approaches for quantitation of hydrocortisone acetate and pramoxine hydrochloride in presence of their impurities
AU - Ibrahim, Fawzia
AU - El-Deen, Asmaa Kamal
AU - Shimizu, Kuniyoshi
N1 - Publisher Copyright:
© 2018
PY - 2018/7
Y1 - 2018/7
N2 - In the present study, we compare the performance of two reversed-phase liquid chromatographic approaches using different eluents either conventional hydro-organic eluent or micellar one for simultaneous estimation of hydrocortisone acetate and pramoxine hydrochloride in presence of their degradants and process-related impurities; hydrocortisone and 4-butoxyphenol, respectively. For conventional reversed-phase liquid chromatography (RPLC), separation of the studied compounds was completed on an Inertsil ODS 3-C18 column (150 mm × 4.6 mm, 5 μm particle size) with a mobile phase consists of 50 mM phosphate buffer (pH 5.0): acetonitrile (50: 50, v/v). For micellar liquid chromatography (MLC), an Eclipse XDB-C8 column (150 mm × 4.6 mm, 5 μm particle size) was chosen for the separation with a green mobile phase consists of 0.15 M sodium dodecyl sulfate, 0.3% triethylamine and 10% n-butanol in 20 mM orthophosphoric acid (pH 5.0). Both methods were extended to analyze hydrocortisone acetate and pramoxine hydrochloride in their co-formulated cream. RPLC was superior to MLC with regard to sensitivity for the estimation of impurities. While, MLC represents an eco-friendly, less hazardous and biodegradable approach. Furthermore, the direct injection of the cream to the system without the need to laborious samples pretreatment, excessive amount of analysis time and/or use of large amount of toxic organic solvents is one of the outstanding advantages of MLC.
AB - In the present study, we compare the performance of two reversed-phase liquid chromatographic approaches using different eluents either conventional hydro-organic eluent or micellar one for simultaneous estimation of hydrocortisone acetate and pramoxine hydrochloride in presence of their degradants and process-related impurities; hydrocortisone and 4-butoxyphenol, respectively. For conventional reversed-phase liquid chromatography (RPLC), separation of the studied compounds was completed on an Inertsil ODS 3-C18 column (150 mm × 4.6 mm, 5 μm particle size) with a mobile phase consists of 50 mM phosphate buffer (pH 5.0): acetonitrile (50: 50, v/v). For micellar liquid chromatography (MLC), an Eclipse XDB-C8 column (150 mm × 4.6 mm, 5 μm particle size) was chosen for the separation with a green mobile phase consists of 0.15 M sodium dodecyl sulfate, 0.3% triethylamine and 10% n-butanol in 20 mM orthophosphoric acid (pH 5.0). Both methods were extended to analyze hydrocortisone acetate and pramoxine hydrochloride in their co-formulated cream. RPLC was superior to MLC with regard to sensitivity for the estimation of impurities. While, MLC represents an eco-friendly, less hazardous and biodegradable approach. Furthermore, the direct injection of the cream to the system without the need to laborious samples pretreatment, excessive amount of analysis time and/or use of large amount of toxic organic solvents is one of the outstanding advantages of MLC.
UR - http://www.scopus.com/inward/record.url?scp=85041179680&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85041179680&partnerID=8YFLogxK
U2 - 10.1016/j.jfda.2017.12.008
DO - 10.1016/j.jfda.2017.12.008
M3 - Article
C2 - 29976408
AN - SCOPUS:85041179680
SN - 1021-9498
VL - 26
SP - 1160
EP - 1170
JO - Journal of Food and Drug Analysis
JF - Journal of Food and Drug Analysis
IS - 3
ER -