Effect of ambient light variations on luminance ratio of CRT monitor and image contrast in ultrasonography

Shiro Hatanaka, Junji Morishita, Akiko Chishaki, Yukari Noguchi, Eisaku Hokazono, Yoko Komiya, Yoshiharu Higashida, Yoshinobu Shimanoe, Kunio Nakamura, Hiroyuki Tashiro

Research output: Contribution to journalArticlepeer-review

Abstract

Variation in the luminance ratio of a cathode ray tube(CRT)monitor and the ultrasonographic images at different levels of ambient light(0-150 lux)was investigated to obtain optimum ambient light in the ultrasonography suite. The maximum and minimum luminances of test patterns and ultrasonographic images were measured after three technicians independently optimized the brightness and contrast of the CRT monitor and ultrasonographic images at different levels of ambient light. Furthermore, the luminance ratio was calculated from the maximum luminance divided by the minimum luminance. When ambient light increased, it was difficult for the technicians to optimize the brightness and contrast settings of the CRT monitor to maintain a high luminance ratio at 0 lux. The luminance ratio decreased rapidly as ambient light increased up to 20 lux. However, the luminance ratio decreased gradually when ambient light was higher than 20 lux. It is necessary to take into consideration the ambient light to maintain a high luminance ratio of ultrasonographic images.

Original languageEnglish
Pages (from-to)1512-1518
Number of pages7
JournalNippon Hoshasen Gijutsu Gakkai zasshi
Volume65
Issue number11
DOIs
Publication statusPublished - Nov 20 2009

All Science Journal Classification (ASJC) codes

  • Medicine(all)

Fingerprint Dive into the research topics of 'Effect of ambient light variations on luminance ratio of CRT monitor and image contrast in ultrasonography'. Together they form a unique fingerprint.

Cite this