Abstract
Objective: The purpose of this study was to examine and compare the four combination of pulse oximeters (POs) and monitoring sites, the Nihon Kohden BSS-9800 (N), the Masimo SET Radical (M), the Nellcor N550 D-25 (N-D) and the Nellcor N550 Max-Fast (N-MF) in patients with peripheral hypoperfusion. Methods: About 20 adult patients undergoing cardiac surgery using mild hypothermic cardiopulmonary bypass (CPB) were studied prospectively. PO sensors were applied on fingers in N, M and N-D, while on the forehead in N-MF. Results: PO failure was defined as failure to show no SpO2 value or incorrect SpO2 values. PO failure occurred in 12 patients with N, ten patients with M, four patients with N-D and ten patients with N-MF, respectively (p < 0.05 N-D vs. N, M, N-MF). The duration of PO failure was 19+/-30% of aortic cross-clamping with N, 29 +/-33% with M, 10 +/-26% with N-D and 43 +/- 57% with N-MF, respectively (p < -0.05 N-D vs. M and N-MF). Conclusions: The results suggested that N-D is most useful among four combinations of POs and monitoring sites tested in this study for monitoring SpO2 during hypoperfusion. The superiority of N-MF during hypoperfusion was not evident in the present study.
Original language | English |
---|---|
Pages (from-to) | 249-252 |
Number of pages | 4 |
Journal | Journal of Clinical Monitoring and Computing |
Volume | 21 |
Issue number | 4 |
DOIs | |
Publication status | Published - Aug 2007 |