T1 - Examination of beam profile measurement using an imaging plate by the light exposure fading method for quality assurance of external radiation therapy.
AU - Fujibuchi, Toshioh
AU - Funabashi, Nobusada
AU - Hashimoto, Masatoshi
AU - Abe, Yukinao
AU - Iimori, Takashi
AU - Matsubayashi, Fumiyasu
AU - Sawada, Kohichi
AU - Kurokawa, Masayuki
AU - Isobe, Tomonori
AU - Kikawa, Takashi
AU - Komuro, Issei
PY - 2006/1/1
Y1 - 2006/1/1
N2 - High-resolution film dosimetry has been used for several decades to check and to measure two-dimensional dose distributions. However, in recent years, the automatic processor has been replaced by the spread of computed radiography, or has been little used hospitals. In this study, we measured the off-center ratio (OCR) of the open field, after an irradiating radiation beam was delivered to the imaging plate (IP) under conditions in which the IP was exposed to a fixed amount of light with fading, and compared these data with the OCR measured by an ionization-chamber dosimeter, which is the standard method used for measuring radiation dose. Profile measurement using IP could be achieved by performing light fading, even at a range of more than 100 MU. Further, by using a metallic filter, we succeeded in demonstrating that the profile measurement of IP in an open irradiation field could approximate the values of those obtained by an ionization chamber dosimeter. This method can serve as a simple, easy-to-use method for evaluating the QA of dose distribution in radiation therapy.
AB - High-resolution film dosimetry has been used for several decades to check and to measure two-dimensional dose distributions. However, in recent years, the automatic processor has been replaced by the spread of computed radiography, or has been little used hospitals. In this study, we measured the off-center ratio (OCR) of the open field, after an irradiating radiation beam was delivered to the imaging plate (IP) under conditions in which the IP was exposed to a fixed amount of light with fading, and compared these data with the OCR measured by an ionization-chamber dosimeter, which is the standard method used for measuring radiation dose. Profile measurement using IP could be achieved by performing light fading, even at a range of more than 100 MU. Further, by using a metallic filter, we succeeded in demonstrating that the profile measurement of IP in an open irradiation field could approximate the values of those obtained by an ionization chamber dosimeter. This method can serve as a simple, easy-to-use method for evaluating the QA of dose distribution in radiation therapy.
UR - http://www.scopus.com/inward/record.url?scp=39049179291&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=39049179291&partnerID=8YFLogxK
U2 - 10.6009/jjrt.62.1697
DO - 10.6009/jjrt.62.1697
M3 - Article
C2 - 17189937
AN - SCOPUS:39049179291
SN - 0369-4305
VL - 62
SP - 1697
EP - 1706
JO - Nippon Hoshasen Gijutsu Gakkai zasshi
JF - Nippon Hoshasen Gijutsu Gakkai zasshi
IS - 12
ER -