Depending on the facility, a radiation protective apron (protector) is used to protect the gonad from radiation exposure in chest radiography. To determine the necessity of using a protector during chest radiography, we measured the effect of the protector on the gonad in this study. First, using a human body phantom, we measured the absorbed dose of the female gonad with and without the protector, using a thermoluminescence dosimeter (TLD), and confirmed its protective effect. Using the protector, the absorbed dose was reduced to 28+/-2% and 39+/-4% for field sizes of 14 x 17 inch and 14 x 14 inch, respectively. Next, we used Monte Carlo simulation and confirmed, not only the validity of the actual measurement values, but also the fact that the influence of radiation on the absorbed dose of the gonad was mostly from scattered radiation from inside the body for the 14 x 17 inch field size, and also from the X-ray tube for the 14 x 14 inch field size. Although a certain protective effect is achieved by using the protector, the radiation dose to the gonad is only a few microGy even without a protector. Thus, the risk of a genetic effect would be as small as 10(-8). Given that acceptable risk is below 10(-6), we conclude the use of a radiation protective apron is not necessary for diagnostic chest radiography.

Original languageEnglish
Pages (from-to)1704-1712
Number of pages9
JournalNippon Hoshasen Gijutsu Gakkai zasshi
Volume60
Issue number12
DOIs
Publication statusPublished - Dec 2004

Fingerprint

Radiation Effects
Gonads
Radiography
Thorax
Radiation
Human Body
X-Rays

All Science Journal Classification (ASJC) codes

  • Medicine(all)

Cite this

Gonad protective effect of radiation protective apron in chest radiography. / Hashimoto, Masatoshi; Kato, Hideyuki; Fujibuchi, Toshiou; Ochi, Shigehiro; Morita, Fuminori.

In: Nippon Hoshasen Gijutsu Gakkai zasshi, Vol. 60, No. 12, 12.2004, p. 1704-1712.

Research output: Contribution to journalArticle

Hashimoto, Masatoshi ; Kato, Hideyuki ; Fujibuchi, Toshiou ; Ochi, Shigehiro ; Morita, Fuminori. / Gonad protective effect of radiation protective apron in chest radiography. In: Nippon Hoshasen Gijutsu Gakkai zasshi. 2004 ; Vol. 60, No. 12. pp. 1704-1712.
@article{27535390621344d6a0f7df1abd95166a,
title = "Gonad protective effect of radiation protective apron in chest radiography",
abstract = "Depending on the facility, a radiation protective apron (protector) is used to protect the gonad from radiation exposure in chest radiography. To determine the necessity of using a protector during chest radiography, we measured the effect of the protector on the gonad in this study. First, using a human body phantom, we measured the absorbed dose of the female gonad with and without the protector, using a thermoluminescence dosimeter (TLD), and confirmed its protective effect. Using the protector, the absorbed dose was reduced to 28+/-2{\%} and 39+/-4{\%} for field sizes of 14 x 17 inch and 14 x 14 inch, respectively. Next, we used Monte Carlo simulation and confirmed, not only the validity of the actual measurement values, but also the fact that the influence of radiation on the absorbed dose of the gonad was mostly from scattered radiation from inside the body for the 14 x 17 inch field size, and also from the X-ray tube for the 14 x 14 inch field size. Although a certain protective effect is achieved by using the protector, the radiation dose to the gonad is only a few microGy even without a protector. Thus, the risk of a genetic effect would be as small as 10(-8). Given that acceptable risk is below 10(-6), we conclude the use of a radiation protective apron is not necessary for diagnostic chest radiography.",
author = "Masatoshi Hashimoto and Hideyuki Kato and Toshiou Fujibuchi and Shigehiro Ochi and Fuminori Morita",
year = "2004",
month = "12",
doi = "10.6009/jjrt.KJ00003560635",
language = "English",
volume = "60",
pages = "1704--1712",
journal = "Nippon Hoshasen Gijutsu Gakkai zasshi",
issn = "0369-4305",
publisher = "Nihon Hoshasen Gijutsu Gakkai/Japanese Society of Radiological Technology",
number = "12",

}

TY - JOUR

T1 - Gonad protective effect of radiation protective apron in chest radiography

AU - Hashimoto, Masatoshi

AU - Kato, Hideyuki

AU - Fujibuchi, Toshiou

AU - Ochi, Shigehiro

AU - Morita, Fuminori

PY - 2004/12

Y1 - 2004/12

N2 - Depending on the facility, a radiation protective apron (protector) is used to protect the gonad from radiation exposure in chest radiography. To determine the necessity of using a protector during chest radiography, we measured the effect of the protector on the gonad in this study. First, using a human body phantom, we measured the absorbed dose of the female gonad with and without the protector, using a thermoluminescence dosimeter (TLD), and confirmed its protective effect. Using the protector, the absorbed dose was reduced to 28+/-2% and 39+/-4% for field sizes of 14 x 17 inch and 14 x 14 inch, respectively. Next, we used Monte Carlo simulation and confirmed, not only the validity of the actual measurement values, but also the fact that the influence of radiation on the absorbed dose of the gonad was mostly from scattered radiation from inside the body for the 14 x 17 inch field size, and also from the X-ray tube for the 14 x 14 inch field size. Although a certain protective effect is achieved by using the protector, the radiation dose to the gonad is only a few microGy even without a protector. Thus, the risk of a genetic effect would be as small as 10(-8). Given that acceptable risk is below 10(-6), we conclude the use of a radiation protective apron is not necessary for diagnostic chest radiography.

AB - Depending on the facility, a radiation protective apron (protector) is used to protect the gonad from radiation exposure in chest radiography. To determine the necessity of using a protector during chest radiography, we measured the effect of the protector on the gonad in this study. First, using a human body phantom, we measured the absorbed dose of the female gonad with and without the protector, using a thermoluminescence dosimeter (TLD), and confirmed its protective effect. Using the protector, the absorbed dose was reduced to 28+/-2% and 39+/-4% for field sizes of 14 x 17 inch and 14 x 14 inch, respectively. Next, we used Monte Carlo simulation and confirmed, not only the validity of the actual measurement values, but also the fact that the influence of radiation on the absorbed dose of the gonad was mostly from scattered radiation from inside the body for the 14 x 17 inch field size, and also from the X-ray tube for the 14 x 14 inch field size. Although a certain protective effect is achieved by using the protector, the radiation dose to the gonad is only a few microGy even without a protector. Thus, the risk of a genetic effect would be as small as 10(-8). Given that acceptable risk is below 10(-6), we conclude the use of a radiation protective apron is not necessary for diagnostic chest radiography.

UR - http://www.scopus.com/inward/record.url?scp=16644385832&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=16644385832&partnerID=8YFLogxK

U2 - 10.6009/jjrt.KJ00003560635

DO - 10.6009/jjrt.KJ00003560635

M3 - Article

C2 - 15614222

AN - SCOPUS:16644385832

VL - 60

SP - 1704

EP - 1712

JO - Nippon Hoshasen Gijutsu Gakkai zasshi

JF - Nippon Hoshasen Gijutsu Gakkai zasshi

SN - 0369-4305

IS - 12

ER -