Abstract
The high performance bench-top cylindrical magnetic shield with magnetic shaking enhancement was discussed. The shielding factors obtained at the center were 10300 for the transverse direction and 520 for the axial direction. It was found that axial shielding factor can be easily increased by adding third shell of axial structure.
Original language | English |
---|---|
Pages (from-to) | EV09 |
Journal | Digests of the Intermag Conference |
Publication status | Published - 2002 |
Event | 2002 IEEE International Magnetics Conference-2002 IEEE INTERMAG - Amsterdam, Netherlands Duration: Apr 28 2002 → May 2 2002 |
All Science Journal Classification (ASJC) codes
- Electrical and Electronic Engineering