IFCC reference procedures for measurement of the catalytic concentrations of enzymes: Corrigendum, notes and useful advice

Gerhard Schumann, Francesca Canalias, Poul J. Joergensen, Dongchon Kang, Jean Marc Lessinger, Rainer Klauke

Research output: Contribution to journalArticle

11 Citations (Scopus)

Abstract

The primary reference measurement procedures (PRMPs) for the international standardization of catalytic concentration measurements of α-amylase, alanine aminotransferase, aspartate aminotransferase (AST), creatine kinase (CK), γ-glutamyltransferase and lactate dehydrogenase have been performed in reference laboratories for several years. The IFCC Committee on Reference Systems for Enzymes and two reference laboratories, with official accreditation for the PRMPs, have collected useful information on some of the steps of the reference procedures that require special attention. This document comprises errata corrige for minor mistakes in published PRMPs for AST and CK. Several notes on the PRMPs are emphasized. This includes details that are very important for improved standardization, and general suggestions for reducing measurement uncertainty.

Original languageEnglish
Pages (from-to)615-621
Number of pages7
JournalClinical Chemistry and Laboratory Medicine
Volume48
Issue number5
DOIs
Publication statusPublished - May 1 2010

All Science Journal Classification (ASJC) codes

  • Clinical Biochemistry
  • Biochemistry, medical

Fingerprint Dive into the research topics of 'IFCC reference procedures for measurement of the catalytic concentrations of enzymes: Corrigendum, notes and useful advice'. Together they form a unique fingerprint.