[Improvement of quantitative accuracy using phase-based respiratory-gated PET/CT in phantom and clinical studies]

Kenta Miwa, Kei Wagatsuma, Takuro Umeda, Noriaki Miyaji, Taisuke Murata, Atsushi Osawa, Tomohiro Takiguchi, Mitsuru Koizumi, Masayuki Sasaki

Research output: Contribution to journalArticle

Abstract

OBJECTIVE: The present study aimed at determining the quantitative accuracy of phase-based respiratory-gated PET/CT imaging using phantom and clinical studies.

METHODS: The effects of target size, target-to-background ratio (TBR), and respiratory motion on PET images were estimated using a NEMA body phantom comprising six spheres (diameter 10-37mm) in a solution of F-18 of three different TBRs (4, 6, 8). The phantom was moved in a superior-inferior direction at motion displacements of 0, 10, 20 and 30 mm. Stationary images of the phantom as well as non-gated (3D) and gated (4D) images of the phantom while moving were reconstructed and the recovery coefficient (RC) of individual spheres was calculated from each image. We then determined the RC improvement rate to evaluate improvements conferred by 4D-PET/CT. We retrospectively analyzed data from 14 patients with lung cancer who were examined by 3D- and 4D-PET/CT. Each lesion on the 3D-PET/CT and each of the five phases of the 4D-PET/CT were analyzed.

RESULTS: Larger motion displacement and TBR resulted in increased RC degradation for small spheres. The RC improvement rate showed that 4D acquisition improved the RC of spheres with larger motion displacement exceeding 13 mm in diameter. 4D-PET/CT alone can reduce the effects of motion blurring, but partial volume effects may still be the dominant source of quantitative inaccuracy for small lesions. The trends of phantom and clinical studies for evaluating the improvement rate were similar.

CONCLUSIONS: 4D-PET/CT significantly improved the quantitative accuracy of PET images particularly when larger motion displacement exceeded 17mm in diameter such as in lung cancer.

Original languageEnglish
Pages (from-to)1235-1242
Number of pages8
JournalNihon Hoshasen Gijutsu Gakkai zasshi
Volume70
Issue number11
DOIs
Publication statusPublished - Nov 1 2014

Fingerprint

Four-Dimensional Computed Tomography
Lung Neoplasms
Imaging Phantoms
Clinical Studies

All Science Journal Classification (ASJC) codes

  • Medicine(all)

Cite this

[Improvement of quantitative accuracy using phase-based respiratory-gated PET/CT in phantom and clinical studies]. / Miwa, Kenta; Wagatsuma, Kei; Umeda, Takuro; Miyaji, Noriaki; Murata, Taisuke; Osawa, Atsushi; Takiguchi, Tomohiro; Koizumi, Mitsuru; Sasaki, Masayuki.

In: Nihon Hoshasen Gijutsu Gakkai zasshi, Vol. 70, No. 11, 01.11.2014, p. 1235-1242.

Research output: Contribution to journalArticle

Miwa, Kenta ; Wagatsuma, Kei ; Umeda, Takuro ; Miyaji, Noriaki ; Murata, Taisuke ; Osawa, Atsushi ; Takiguchi, Tomohiro ; Koizumi, Mitsuru ; Sasaki, Masayuki. / [Improvement of quantitative accuracy using phase-based respiratory-gated PET/CT in phantom and clinical studies]. In: Nihon Hoshasen Gijutsu Gakkai zasshi. 2014 ; Vol. 70, No. 11. pp. 1235-1242.
@article{a634271dbc01400db0e99c95ae98041e,
title = "[Improvement of quantitative accuracy using phase-based respiratory-gated PET/CT in phantom and clinical studies]",
abstract = "OBJECTIVE: The present study aimed at determining the quantitative accuracy of phase-based respiratory-gated PET/CT imaging using phantom and clinical studies.METHODS: The effects of target size, target-to-background ratio (TBR), and respiratory motion on PET images were estimated using a NEMA body phantom comprising six spheres (diameter 10-37mm) in a solution of F-18 of three different TBRs (4, 6, 8). The phantom was moved in a superior-inferior direction at motion displacements of 0, 10, 20 and 30 mm. Stationary images of the phantom as well as non-gated (3D) and gated (4D) images of the phantom while moving were reconstructed and the recovery coefficient (RC) of individual spheres was calculated from each image. We then determined the RC improvement rate to evaluate improvements conferred by 4D-PET/CT. We retrospectively analyzed data from 14 patients with lung cancer who were examined by 3D- and 4D-PET/CT. Each lesion on the 3D-PET/CT and each of the five phases of the 4D-PET/CT were analyzed.RESULTS: Larger motion displacement and TBR resulted in increased RC degradation for small spheres. The RC improvement rate showed that 4D acquisition improved the RC of spheres with larger motion displacement exceeding 13 mm in diameter. 4D-PET/CT alone can reduce the effects of motion blurring, but partial volume effects may still be the dominant source of quantitative inaccuracy for small lesions. The trends of phantom and clinical studies for evaluating the improvement rate were similar.CONCLUSIONS: 4D-PET/CT significantly improved the quantitative accuracy of PET images particularly when larger motion displacement exceeded 17mm in diameter such as in lung cancer.",
author = "Kenta Miwa and Kei Wagatsuma and Takuro Umeda and Noriaki Miyaji and Taisuke Murata and Atsushi Osawa and Tomohiro Takiguchi and Mitsuru Koizumi and Masayuki Sasaki",
year = "2014",
month = "11",
day = "1",
doi = "10.6009/jjrt.2014_JSRT_70.11.1235",
language = "English",
volume = "70",
pages = "1235--1242",
journal = "Nippon Hoshasen Gijutsu Gakkai zasshi",
issn = "0369-4305",
publisher = "Nihon Hoshasen Gijutsu Gakkai/Japanese Society of Radiological Technology",
number = "11",

}

TY - JOUR

T1 - [Improvement of quantitative accuracy using phase-based respiratory-gated PET/CT in phantom and clinical studies]

AU - Miwa, Kenta

AU - Wagatsuma, Kei

AU - Umeda, Takuro

AU - Miyaji, Noriaki

AU - Murata, Taisuke

AU - Osawa, Atsushi

AU - Takiguchi, Tomohiro

AU - Koizumi, Mitsuru

AU - Sasaki, Masayuki

PY - 2014/11/1

Y1 - 2014/11/1

N2 - OBJECTIVE: The present study aimed at determining the quantitative accuracy of phase-based respiratory-gated PET/CT imaging using phantom and clinical studies.METHODS: The effects of target size, target-to-background ratio (TBR), and respiratory motion on PET images were estimated using a NEMA body phantom comprising six spheres (diameter 10-37mm) in a solution of F-18 of three different TBRs (4, 6, 8). The phantom was moved in a superior-inferior direction at motion displacements of 0, 10, 20 and 30 mm. Stationary images of the phantom as well as non-gated (3D) and gated (4D) images of the phantom while moving were reconstructed and the recovery coefficient (RC) of individual spheres was calculated from each image. We then determined the RC improvement rate to evaluate improvements conferred by 4D-PET/CT. We retrospectively analyzed data from 14 patients with lung cancer who were examined by 3D- and 4D-PET/CT. Each lesion on the 3D-PET/CT and each of the five phases of the 4D-PET/CT were analyzed.RESULTS: Larger motion displacement and TBR resulted in increased RC degradation for small spheres. The RC improvement rate showed that 4D acquisition improved the RC of spheres with larger motion displacement exceeding 13 mm in diameter. 4D-PET/CT alone can reduce the effects of motion blurring, but partial volume effects may still be the dominant source of quantitative inaccuracy for small lesions. The trends of phantom and clinical studies for evaluating the improvement rate were similar.CONCLUSIONS: 4D-PET/CT significantly improved the quantitative accuracy of PET images particularly when larger motion displacement exceeded 17mm in diameter such as in lung cancer.

AB - OBJECTIVE: The present study aimed at determining the quantitative accuracy of phase-based respiratory-gated PET/CT imaging using phantom and clinical studies.METHODS: The effects of target size, target-to-background ratio (TBR), and respiratory motion on PET images were estimated using a NEMA body phantom comprising six spheres (diameter 10-37mm) in a solution of F-18 of three different TBRs (4, 6, 8). The phantom was moved in a superior-inferior direction at motion displacements of 0, 10, 20 and 30 mm. Stationary images of the phantom as well as non-gated (3D) and gated (4D) images of the phantom while moving were reconstructed and the recovery coefficient (RC) of individual spheres was calculated from each image. We then determined the RC improvement rate to evaluate improvements conferred by 4D-PET/CT. We retrospectively analyzed data from 14 patients with lung cancer who were examined by 3D- and 4D-PET/CT. Each lesion on the 3D-PET/CT and each of the five phases of the 4D-PET/CT were analyzed.RESULTS: Larger motion displacement and TBR resulted in increased RC degradation for small spheres. The RC improvement rate showed that 4D acquisition improved the RC of spheres with larger motion displacement exceeding 13 mm in diameter. 4D-PET/CT alone can reduce the effects of motion blurring, but partial volume effects may still be the dominant source of quantitative inaccuracy for small lesions. The trends of phantom and clinical studies for evaluating the improvement rate were similar.CONCLUSIONS: 4D-PET/CT significantly improved the quantitative accuracy of PET images particularly when larger motion displacement exceeded 17mm in diameter such as in lung cancer.

UR - http://www.scopus.com/inward/record.url?scp=84922481281&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84922481281&partnerID=8YFLogxK

U2 - 10.6009/jjrt.2014_JSRT_70.11.1235

DO - 10.6009/jjrt.2014_JSRT_70.11.1235

M3 - Article

C2 - 25410329

AN - SCOPUS:84922481281

VL - 70

SP - 1235

EP - 1242

JO - Nippon Hoshasen Gijutsu Gakkai zasshi

JF - Nippon Hoshasen Gijutsu Gakkai zasshi

SN - 0369-4305

IS - 11

ER -