Influence of light output calibration on neutron energy spectrum unfolding up to 300 MeV using liquid organic scintillator

Eunji Lee, Nobuhiro Shigyo, Tsuyoshi Kajimoto, Toshiya Sanami, Naruhiro Matsufuji, Shogo Izumitani, Naoki Tokunaga, Mamoru Kiyota

Research output: Contribution to journalArticle

Abstract

The influence from uncertainty on light output calibration was studied experimentally for neutron energy spectra up to 300 MeV obtained by an unfolding method using a liquid organic scintillator. The neutrons were generated from the interactions of 100 and 290 MeV/u 28 Si ion beams on thick silicon targets at a 75° direction with respect to the beam axis. The light output calibration determined by using Compton edges of γ-rays from 60 Co and 241 Am-Be sources was used for the unfolding method. The uncertainty of the calibration points at the Compton edges was estimated to be within 3%. Two calibration lines which connected the 3% larger light output point of the Compton edge of the γ-ray from 60 Co and that the 3% lower light output point of that from 241 Am-Be and vice versa were examined for influence on variation of calibration line. The unfolded spectra using the two calibration lines were compared with one using the calibration line connecting light output points of Compton edges from both γ-ray sources. The comparison indicates the uncertainty of calibration line influences the neutron light output spectrum within order of several %. The all unfolded spectra generally reproduced one by the time-of-flight (TOF) method in the same experiment. The difference between TOF and unfolded spectra were 17 and 8% for 100 and 290 MeV/u data, at maximum.

Original languageEnglish
Pages (from-to)26-33
Number of pages8
JournalNuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms
Volume445
DOIs
Publication statusPublished - Apr 15 2019

Fingerprint

organic liquids
neutron spectra
Phosphors
scintillation counters
Neutrons
energy spectra
Calibration
output
Liquids
rays
neutrons
ion beams
Ion beams
silicon
Silicon
interactions

All Science Journal Classification (ASJC) codes

  • Nuclear and High Energy Physics
  • Instrumentation

Cite this

Influence of light output calibration on neutron energy spectrum unfolding up to 300 MeV using liquid organic scintillator. / Lee, Eunji; Shigyo, Nobuhiro; Kajimoto, Tsuyoshi; Sanami, Toshiya; Matsufuji, Naruhiro; Izumitani, Shogo; Tokunaga, Naoki; Kiyota, Mamoru.

In: Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms, Vol. 445, 15.04.2019, p. 26-33.

Research output: Contribution to journalArticle

@article{4cbc312cad6d4308aec2f9eed5532bc1,
title = "Influence of light output calibration on neutron energy spectrum unfolding up to 300 MeV using liquid organic scintillator",
abstract = "The influence from uncertainty on light output calibration was studied experimentally for neutron energy spectra up to 300 MeV obtained by an unfolding method using a liquid organic scintillator. The neutrons were generated from the interactions of 100 and 290 MeV/u 28 Si ion beams on thick silicon targets at a 75° direction with respect to the beam axis. The light output calibration determined by using Compton edges of γ-rays from 60 Co and 241 Am-Be sources was used for the unfolding method. The uncertainty of the calibration points at the Compton edges was estimated to be within 3{\%}. Two calibration lines which connected the 3{\%} larger light output point of the Compton edge of the γ-ray from 60 Co and that the 3{\%} lower light output point of that from 241 Am-Be and vice versa were examined for influence on variation of calibration line. The unfolded spectra using the two calibration lines were compared with one using the calibration line connecting light output points of Compton edges from both γ-ray sources. The comparison indicates the uncertainty of calibration line influences the neutron light output spectrum within order of several {\%}. The all unfolded spectra generally reproduced one by the time-of-flight (TOF) method in the same experiment. The difference between TOF and unfolded spectra were 17 and 8{\%} for 100 and 290 MeV/u data, at maximum.",
author = "Eunji Lee and Nobuhiro Shigyo and Tsuyoshi Kajimoto and Toshiya Sanami and Naruhiro Matsufuji and Shogo Izumitani and Naoki Tokunaga and Mamoru Kiyota",
year = "2019",
month = "4",
day = "15",
doi = "10.1016/j.nimb.2019.02.024",
language = "English",
volume = "445",
pages = "26--33",
journal = "Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms",
issn = "0168-583X",
publisher = "Elsevier",

}

TY - JOUR

T1 - Influence of light output calibration on neutron energy spectrum unfolding up to 300 MeV using liquid organic scintillator

AU - Lee, Eunji

AU - Shigyo, Nobuhiro

AU - Kajimoto, Tsuyoshi

AU - Sanami, Toshiya

AU - Matsufuji, Naruhiro

AU - Izumitani, Shogo

AU - Tokunaga, Naoki

AU - Kiyota, Mamoru

PY - 2019/4/15

Y1 - 2019/4/15

N2 - The influence from uncertainty on light output calibration was studied experimentally for neutron energy spectra up to 300 MeV obtained by an unfolding method using a liquid organic scintillator. The neutrons were generated from the interactions of 100 and 290 MeV/u 28 Si ion beams on thick silicon targets at a 75° direction with respect to the beam axis. The light output calibration determined by using Compton edges of γ-rays from 60 Co and 241 Am-Be sources was used for the unfolding method. The uncertainty of the calibration points at the Compton edges was estimated to be within 3%. Two calibration lines which connected the 3% larger light output point of the Compton edge of the γ-ray from 60 Co and that the 3% lower light output point of that from 241 Am-Be and vice versa were examined for influence on variation of calibration line. The unfolded spectra using the two calibration lines were compared with one using the calibration line connecting light output points of Compton edges from both γ-ray sources. The comparison indicates the uncertainty of calibration line influences the neutron light output spectrum within order of several %. The all unfolded spectra generally reproduced one by the time-of-flight (TOF) method in the same experiment. The difference between TOF and unfolded spectra were 17 and 8% for 100 and 290 MeV/u data, at maximum.

AB - The influence from uncertainty on light output calibration was studied experimentally for neutron energy spectra up to 300 MeV obtained by an unfolding method using a liquid organic scintillator. The neutrons were generated from the interactions of 100 and 290 MeV/u 28 Si ion beams on thick silicon targets at a 75° direction with respect to the beam axis. The light output calibration determined by using Compton edges of γ-rays from 60 Co and 241 Am-Be sources was used for the unfolding method. The uncertainty of the calibration points at the Compton edges was estimated to be within 3%. Two calibration lines which connected the 3% larger light output point of the Compton edge of the γ-ray from 60 Co and that the 3% lower light output point of that from 241 Am-Be and vice versa were examined for influence on variation of calibration line. The unfolded spectra using the two calibration lines were compared with one using the calibration line connecting light output points of Compton edges from both γ-ray sources. The comparison indicates the uncertainty of calibration line influences the neutron light output spectrum within order of several %. The all unfolded spectra generally reproduced one by the time-of-flight (TOF) method in the same experiment. The difference between TOF and unfolded spectra were 17 and 8% for 100 and 290 MeV/u data, at maximum.

UR - http://www.scopus.com/inward/record.url?scp=85062374350&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85062374350&partnerID=8YFLogxK

U2 - 10.1016/j.nimb.2019.02.024

DO - 10.1016/j.nimb.2019.02.024

M3 - Article

AN - SCOPUS:85062374350

VL - 445

SP - 26

EP - 33

JO - Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms

JF - Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms

SN - 0168-583X

ER -