Influence of myopotential interference on the Wavelet discrimination algorithm in implantable cardioverter-defibrillator

Kazuya Mizukami, Hisashi Yokoshiki, Hirofumi Mitsuyama, Masaya Watanabe, Taro Tenma, Rui Kamada, Masayuki Takahashi, Ryo Sasaki, Motoki Maeno, Hiroyuki Tsutsui

Research output: Contribution to journalArticlepeer-review

Abstract

Background Wavelet is a morphology-based algorithm for detecting ventricular tachycardia. The electrogram (EGM) source of the Wavelet algorithm is nominally programmed with the Can-RV coil configuration, which records a far-field ventricular potential. Therefore, it may be influenced by myopotential interference. Methods We performed a retrospective review of 40 outpatients who had an implantable cardioverter-defibrillator (ICD) with the Wavelet algorithm. The percent-match score of the Wavelet algorithm was measured during the isometric chest press by pressing the palms together. We classified patients with percent-match scores below 70% due to myopotential interference as positive morphology change, and those with 70% or more as negative morphology change. Stored episodes of tachycardia were evaluated during the follow-up. Results The number of patients in the positive morphology change group was 22 (55%). Amplitude of the Can-RV coil EGM was lower in the positive morphology change group compared to that in the negative group (3.9±1.3 mV vs. 7.4±1.6 mV, P=0.0015). The cut-off value of the Can-RV coil EGM was 5 mV (area under curve, 0.89). Inappropriate detections caused by myopotential interference occurred in two patients (5%) during a mean follow-up period of 49 months, and one of them received an inappropriate ICD shock. These patients had exhibited positive morphology change. Conclusions The Wavelet algorithm is influenced by myopotential interference when the Can-RV coil EGM is less than 5 mV.

Original languageEnglish
Pages (from-to)214-219
Number of pages6
Journaljournal of arrhythmia
Volume33
Issue number3
DOIs
Publication statusPublished - Jun 2017
Externally publishedYes

All Science Journal Classification (ASJC) codes

  • Cardiology and Cardiovascular Medicine

Fingerprint Dive into the research topics of 'Influence of myopotential interference on the Wavelet discrimination algorithm in implantable cardioverter-defibrillator'. Together they form a unique fingerprint.

Cite this