T1 - Laparoscopic ultrasound manipulator with a spring-based elastic mechanism
AU - Arata, Jumpei
AU - Fukami, Kazunari
AU - Oguri, Susumu
AU - Onogi, Shinya
AU - Ikeda, Tetsuo
AU - Nakadate, Ryu
AU - Sakaguchi, Masamichi
AU - Akahoshi, Tomohiko
AU - Harada, Kanako
AU - Mitsuishi, Mamoru
AU - Hashizume, Makoto
N1 - Funding Information:
Acknowledgements This research was supported by Japan Agency for Medical Research and Development (AMED), Research on Development of New Medical Devices (15hk0102005h0002).
Publisher Copyright:
© 2018, CARS.
PY - 2018/7/1
Y1 - 2018/7/1
N2 - Purpose: Image guidance is a key technology that can improve the outcome of laparoscopic surgery. However, due to the large deformation caused by digestive organs, a computer-aided navigation system based on preoperative imaging data cannot indicate the correct target position of the lesion (e.g., liver tumors and vessels invisible from the organ surface). To overcome this issue, we developed a laparoscopic ultrasound manipulator with two motorized degrees of freedom at the tip, allowing for the performance of a dexterous ultrasound scan in a confined laparoscopic surgical area. Method: The developed manipulator consists of a compact and elastic structure using springs, enabling a safe ultrasound scan and avoiding excess force on the inspected organs. The manipulator is a handheld device equipped with four buttons at the handle, which the surgeon directly grasps to send a motion command to the tip structure. The developed prototype realizes two motorized degree-of-freedom motion at the tip. The size of prototype is 15.0 mm in diameter that is usable in conventional laparoscopy. The tip of the manipulator was carefully designed by considering the kinematic model and the results of the finite element analysis. Results: To assess the prototype, accuracy and rigidity were measured by using a motion processing microscope. The accuracy test showed that the proposed device has a fairly accurate characteristic as a handheld device. This was supposedly caused by the nature of compliant mechanism, which does not have mechanical play in motion. In addition, the intrinsic elastic structure (approximately 2.0 N/mm in most of the range of motion) allowed the ultrasound probe to adequately fit on the curved organ surface without extra effort of manipulation during the inspection. In the in vivo experiment, the yaw motion was found to be effective for investigating the vascular network because the manipulator allows the probe to be rotated while maintaining the same position. Conclusion: The mechanical evaluation and in vivo test results showed high feasibility of the prototype. We are currently working on further mechanical improvement for commercialization and development of a real-time navigation system that can perform three-dimensional reconstruction of ultrasonographic images by implementing a magnetic position sensor at the tip of the manipulator.
AB - Purpose: Image guidance is a key technology that can improve the outcome of laparoscopic surgery. However, due to the large deformation caused by digestive organs, a computer-aided navigation system based on preoperative imaging data cannot indicate the correct target position of the lesion (e.g., liver tumors and vessels invisible from the organ surface). To overcome this issue, we developed a laparoscopic ultrasound manipulator with two motorized degrees of freedom at the tip, allowing for the performance of a dexterous ultrasound scan in a confined laparoscopic surgical area. Method: The developed manipulator consists of a compact and elastic structure using springs, enabling a safe ultrasound scan and avoiding excess force on the inspected organs. The manipulator is a handheld device equipped with four buttons at the handle, which the surgeon directly grasps to send a motion command to the tip structure. The developed prototype realizes two motorized degree-of-freedom motion at the tip. The size of prototype is 15.0 mm in diameter that is usable in conventional laparoscopy. The tip of the manipulator was carefully designed by considering the kinematic model and the results of the finite element analysis. Results: To assess the prototype, accuracy and rigidity were measured by using a motion processing microscope. The accuracy test showed that the proposed device has a fairly accurate characteristic as a handheld device. This was supposedly caused by the nature of compliant mechanism, which does not have mechanical play in motion. In addition, the intrinsic elastic structure (approximately 2.0 N/mm in most of the range of motion) allowed the ultrasound probe to adequately fit on the curved organ surface without extra effort of manipulation during the inspection. In the in vivo experiment, the yaw motion was found to be effective for investigating the vascular network because the manipulator allows the probe to be rotated while maintaining the same position. Conclusion: The mechanical evaluation and in vivo test results showed high feasibility of the prototype. We are currently working on further mechanical improvement for commercialization and development of a real-time navigation system that can perform three-dimensional reconstruction of ultrasonographic images by implementing a magnetic position sensor at the tip of the manipulator.
UR - http://www.scopus.com/inward/record.url?scp=85042629346&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85042629346&partnerID=8YFLogxK
U2 - 10.1007/s11548-018-1709-7
DO - 10.1007/s11548-018-1709-7
M3 - Article
C2 - 29492881
AN - SCOPUS:85042629346
VL - 13
SP - 1063
EP - 1072
JO - Computer-Assisted Radiology and Surgery
JF - Computer-Assisted Radiology and Surgery
SN - 1861-6410
IS - 7
ER -