Liquid-phase immunoassays using brownian relaxation of magnetic markers

Anwarul Kabir Bhuiya, Tetsu Mitake, Masaki Asai, Tomoya Ito, Schunichi Chosakabe, Takashi Yoshida, Keiji Enpuku, Akihiko Kandori

Research output: Contribution to journalArticle

7 Citations (Scopus)

Abstract

We have developed a liquid-phase immunoassay technique using Brownian relaxation of magnetic markers. In this method, biological targets are fixed on the surface of large polymer beads whose size is typically a few μm. When the markers are bound to the targets, their Brownian relaxation time is dominated by that of the polymer bead, becoming much longer than that of unbound (free) markers. The resulting difference between the magnetic properties of the bound and free markers was detected by relaxation measurements. Therefore, we can magnetically distinguish between the bound and free markers, i.e., we can omit a time consuming washing process called bound/free separation. We developed a detection system using a magneto-resistive (MR) sensor and showed that we can detect 1.4× 107 bound markers in 60 μl of solution. If we assume that a single marker is bound to a single target, this sensitivity can be expressed as 3.8× 10-16mol/ml (or 0.38 fmol/ml) in terms of the molecular-number concentration. We also demonstrated the detection of biological targets called biotins, which were conjugated on the surface of the polystyrene beads with a diameter of 3.3 μm. A strong relationship was obtained between the number of bound markers and the number of biotin-conjugated polymer beads, which confirmed the validity of the present method.

Original languageEnglish
Article number6027621
Pages (from-to)2867-2870
Number of pages4
JournalIEEE Transactions on Magnetics
Volume47
Issue number10
DOIs
Publication statusPublished - Oct 1 2011

Fingerprint

Biotin
Polymers
Polystyrenes
Conjugated polymers
Liquids
Washing
Relaxation time
Magnetic properties
Sensors

All Science Journal Classification (ASJC) codes

  • Electrical and Electronic Engineering
  • Electronic, Optical and Magnetic Materials

Cite this

Bhuiya, A. K., Mitake, T., Asai, M., Ito, T., Chosakabe, S., Yoshida, T., ... Kandori, A. (2011). Liquid-phase immunoassays using brownian relaxation of magnetic markers. IEEE Transactions on Magnetics, 47(10), 2867-2870. [6027621]. https://doi.org/10.1109/TMAG.2011.2153828

Liquid-phase immunoassays using brownian relaxation of magnetic markers. / Bhuiya, Anwarul Kabir; Mitake, Tetsu; Asai, Masaki; Ito, Tomoya; Chosakabe, Schunichi; Yoshida, Takashi; Enpuku, Keiji; Kandori, Akihiko.

In: IEEE Transactions on Magnetics, Vol. 47, No. 10, 6027621, 01.10.2011, p. 2867-2870.

Research output: Contribution to journalArticle

Bhuiya, AK, Mitake, T, Asai, M, Ito, T, Chosakabe, S, Yoshida, T, Enpuku, K & Kandori, A 2011, 'Liquid-phase immunoassays using brownian relaxation of magnetic markers', IEEE Transactions on Magnetics, vol. 47, no. 10, 6027621, pp. 2867-2870. https://doi.org/10.1109/TMAG.2011.2153828
Bhuiya, Anwarul Kabir ; Mitake, Tetsu ; Asai, Masaki ; Ito, Tomoya ; Chosakabe, Schunichi ; Yoshida, Takashi ; Enpuku, Keiji ; Kandori, Akihiko. / Liquid-phase immunoassays using brownian relaxation of magnetic markers. In: IEEE Transactions on Magnetics. 2011 ; Vol. 47, No. 10. pp. 2867-2870.
@article{1b5c2ad5f6244eafa9106e1142a4bba5,
title = "Liquid-phase immunoassays using brownian relaxation of magnetic markers",
abstract = "We have developed a liquid-phase immunoassay technique using Brownian relaxation of magnetic markers. In this method, biological targets are fixed on the surface of large polymer beads whose size is typically a few μm. When the markers are bound to the targets, their Brownian relaxation time is dominated by that of the polymer bead, becoming much longer than that of unbound (free) markers. The resulting difference between the magnetic properties of the bound and free markers was detected by relaxation measurements. Therefore, we can magnetically distinguish between the bound and free markers, i.e., we can omit a time consuming washing process called bound/free separation. We developed a detection system using a magneto-resistive (MR) sensor and showed that we can detect 1.4× 107 bound markers in 60 μl of solution. If we assume that a single marker is bound to a single target, this sensitivity can be expressed as 3.8× 10-16mol/ml (or 0.38 fmol/ml) in terms of the molecular-number concentration. We also demonstrated the detection of biological targets called biotins, which were conjugated on the surface of the polystyrene beads with a diameter of 3.3 μm. A strong relationship was obtained between the number of bound markers and the number of biotin-conjugated polymer beads, which confirmed the validity of the present method.",
author = "Bhuiya, {Anwarul Kabir} and Tetsu Mitake and Masaki Asai and Tomoya Ito and Schunichi Chosakabe and Takashi Yoshida and Keiji Enpuku and Akihiko Kandori",
year = "2011",
month = "10",
day = "1",
doi = "10.1109/TMAG.2011.2153828",
language = "English",
volume = "47",
pages = "2867--2870",
journal = "IEEE Transactions on Magnetics",
issn = "0018-9464",
publisher = "Institute of Electrical and Electronics Engineers Inc.",
number = "10",

}

TY - JOUR

T1 - Liquid-phase immunoassays using brownian relaxation of magnetic markers

AU - Bhuiya, Anwarul Kabir

AU - Mitake, Tetsu

AU - Asai, Masaki

AU - Ito, Tomoya

AU - Chosakabe, Schunichi

AU - Yoshida, Takashi

AU - Enpuku, Keiji

AU - Kandori, Akihiko

PY - 2011/10/1

Y1 - 2011/10/1

N2 - We have developed a liquid-phase immunoassay technique using Brownian relaxation of magnetic markers. In this method, biological targets are fixed on the surface of large polymer beads whose size is typically a few μm. When the markers are bound to the targets, their Brownian relaxation time is dominated by that of the polymer bead, becoming much longer than that of unbound (free) markers. The resulting difference between the magnetic properties of the bound and free markers was detected by relaxation measurements. Therefore, we can magnetically distinguish between the bound and free markers, i.e., we can omit a time consuming washing process called bound/free separation. We developed a detection system using a magneto-resistive (MR) sensor and showed that we can detect 1.4× 107 bound markers in 60 μl of solution. If we assume that a single marker is bound to a single target, this sensitivity can be expressed as 3.8× 10-16mol/ml (or 0.38 fmol/ml) in terms of the molecular-number concentration. We also demonstrated the detection of biological targets called biotins, which were conjugated on the surface of the polystyrene beads with a diameter of 3.3 μm. A strong relationship was obtained between the number of bound markers and the number of biotin-conjugated polymer beads, which confirmed the validity of the present method.

AB - We have developed a liquid-phase immunoassay technique using Brownian relaxation of magnetic markers. In this method, biological targets are fixed on the surface of large polymer beads whose size is typically a few μm. When the markers are bound to the targets, their Brownian relaxation time is dominated by that of the polymer bead, becoming much longer than that of unbound (free) markers. The resulting difference between the magnetic properties of the bound and free markers was detected by relaxation measurements. Therefore, we can magnetically distinguish between the bound and free markers, i.e., we can omit a time consuming washing process called bound/free separation. We developed a detection system using a magneto-resistive (MR) sensor and showed that we can detect 1.4× 107 bound markers in 60 μl of solution. If we assume that a single marker is bound to a single target, this sensitivity can be expressed as 3.8× 10-16mol/ml (or 0.38 fmol/ml) in terms of the molecular-number concentration. We also demonstrated the detection of biological targets called biotins, which were conjugated on the surface of the polystyrene beads with a diameter of 3.3 μm. A strong relationship was obtained between the number of bound markers and the number of biotin-conjugated polymer beads, which confirmed the validity of the present method.

UR - http://www.scopus.com/inward/record.url?scp=80053476282&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=80053476282&partnerID=8YFLogxK

U2 - 10.1109/TMAG.2011.2153828

DO - 10.1109/TMAG.2011.2153828

M3 - Article

VL - 47

SP - 2867

EP - 2870

JO - IEEE Transactions on Magnetics

JF - IEEE Transactions on Magnetics

SN - 0018-9464

IS - 10

M1 - 6027621

ER -