Abstract
The triple charge division (TCD) centroid finding method, which uses modified pattern of Backgammon shape cathode (MBSC), is introduced for medium range length position sensitive detectors with optimum numbers of cathode segments. The Graded Charge Division (GCD) employed as a special readout method for TCD is a combination of the charge division readout and sequence grading of serial segments. Linearity prediction of TCD centroid finding and experimental results are used to find an optimum truncation of the apices of MBCS pattern central area.
Original language | English |
---|---|
Pages | 544-547 |
Number of pages | 4 |
Publication status | Published - Dec 1 1996 |
Event | Proceedings of the 1996 IEEE Nuclear Science Symposium. Part 1 (of 3) - Anaheim, CA, USA Duration: Nov 2 1996 → Nov 9 1996 |
Other
Other | Proceedings of the 1996 IEEE Nuclear Science Symposium. Part 1 (of 3) |
---|---|
City | Anaheim, CA, USA |
Period | 11/2/96 → 11/9/96 |
All Science Journal Classification (ASJC) codes
- Computer Vision and Pattern Recognition
- Industrial and Manufacturing Engineering