Necessity of delivering the calibration coefficient of a dosimeter for low energy X-rays used in mammography X-ray units

Chihiro Matsuo, Masayuki Tachibana, Akasaka Tsutomu

Research output: Contribution to journalArticle

Abstract

The calibration coefficient currently supplied by the National Institute of Advanced Industrial Science and Technology (AIST) is measured with an X-ray unit equipped with a W target/Al filter. We compared the calibration coefficient, which is measured with the X-ray unit equipped with a W target/Al filter and the X-ray unit for mammography. For the calibration coefficient, which is measured at the calibration fields of the X-ray units for mammography, the difference in the calibration coefficient measured with the X-ray unit equipped with a W target/Al filter that assumed a difference in the quality index-value of less than 1.5% was 24.2% at the maximum. In addition, many differences of more than unsureness 5% of the calibration coefficient were accepted. Thus, the correct dosimetry is not being performed for mammography. To improve the precision of dosimetry for mammography, at least, it is desirable to employ the calibration coefficient using the Mo target supplied by AIST. Furthermore, the indication of quality of radiation should be made severe, too. To perform equal dosimetry with a diagnosis domain, it is necessary to supply the calibration coefficient of the other target/filters.

Original languageEnglish
Pages (from-to)915-921
Number of pages7
JournalNippon Hoshasen Gijutsu Gakkai zasshi
Volume64
Issue number8
DOIs
Publication statusPublished - Jan 1 2008

Fingerprint

Mammography
Calibration
X-Rays
Technology
Radiation Dosimeters
Radiation

All Science Journal Classification (ASJC) codes

  • Medicine(all)

Cite this

Necessity of delivering the calibration coefficient of a dosimeter for low energy X-rays used in mammography X-ray units. / Matsuo, Chihiro; Tachibana, Masayuki; Tsutomu, Akasaka.

In: Nippon Hoshasen Gijutsu Gakkai zasshi, Vol. 64, No. 8, 01.01.2008, p. 915-921.

Research output: Contribution to journalArticle

@article{36bfc42d5d8745d7ab0b1e78551d88b9,
title = "Necessity of delivering the calibration coefficient of a dosimeter for low energy X-rays used in mammography X-ray units",
abstract = "The calibration coefficient currently supplied by the National Institute of Advanced Industrial Science and Technology (AIST) is measured with an X-ray unit equipped with a W target/Al filter. We compared the calibration coefficient, which is measured with the X-ray unit equipped with a W target/Al filter and the X-ray unit for mammography. For the calibration coefficient, which is measured at the calibration fields of the X-ray units for mammography, the difference in the calibration coefficient measured with the X-ray unit equipped with a W target/Al filter that assumed a difference in the quality index-value of less than 1.5{\%} was 24.2{\%} at the maximum. In addition, many differences of more than unsureness 5{\%} of the calibration coefficient were accepted. Thus, the correct dosimetry is not being performed for mammography. To improve the precision of dosimetry for mammography, at least, it is desirable to employ the calibration coefficient using the Mo target supplied by AIST. Furthermore, the indication of quality of radiation should be made severe, too. To perform equal dosimetry with a diagnosis domain, it is necessary to supply the calibration coefficient of the other target/filters.",
author = "Chihiro Matsuo and Masayuki Tachibana and Akasaka Tsutomu",
year = "2008",
month = "1",
day = "1",
doi = "10.6009/jjrt.64.915",
language = "English",
volume = "64",
pages = "915--921",
journal = "Nippon Hoshasen Gijutsu Gakkai zasshi",
issn = "0369-4305",
publisher = "Nihon Hoshasen Gijutsu Gakkai/Japanese Society of Radiological Technology",
number = "8",

}

TY - JOUR

T1 - Necessity of delivering the calibration coefficient of a dosimeter for low energy X-rays used in mammography X-ray units

AU - Matsuo, Chihiro

AU - Tachibana, Masayuki

AU - Tsutomu, Akasaka

PY - 2008/1/1

Y1 - 2008/1/1

N2 - The calibration coefficient currently supplied by the National Institute of Advanced Industrial Science and Technology (AIST) is measured with an X-ray unit equipped with a W target/Al filter. We compared the calibration coefficient, which is measured with the X-ray unit equipped with a W target/Al filter and the X-ray unit for mammography. For the calibration coefficient, which is measured at the calibration fields of the X-ray units for mammography, the difference in the calibration coefficient measured with the X-ray unit equipped with a W target/Al filter that assumed a difference in the quality index-value of less than 1.5% was 24.2% at the maximum. In addition, many differences of more than unsureness 5% of the calibration coefficient were accepted. Thus, the correct dosimetry is not being performed for mammography. To improve the precision of dosimetry for mammography, at least, it is desirable to employ the calibration coefficient using the Mo target supplied by AIST. Furthermore, the indication of quality of radiation should be made severe, too. To perform equal dosimetry with a diagnosis domain, it is necessary to supply the calibration coefficient of the other target/filters.

AB - The calibration coefficient currently supplied by the National Institute of Advanced Industrial Science and Technology (AIST) is measured with an X-ray unit equipped with a W target/Al filter. We compared the calibration coefficient, which is measured with the X-ray unit equipped with a W target/Al filter and the X-ray unit for mammography. For the calibration coefficient, which is measured at the calibration fields of the X-ray units for mammography, the difference in the calibration coefficient measured with the X-ray unit equipped with a W target/Al filter that assumed a difference in the quality index-value of less than 1.5% was 24.2% at the maximum. In addition, many differences of more than unsureness 5% of the calibration coefficient were accepted. Thus, the correct dosimetry is not being performed for mammography. To improve the precision of dosimetry for mammography, at least, it is desirable to employ the calibration coefficient using the Mo target supplied by AIST. Furthermore, the indication of quality of radiation should be made severe, too. To perform equal dosimetry with a diagnosis domain, it is necessary to supply the calibration coefficient of the other target/filters.

UR - http://www.scopus.com/inward/record.url?scp=54049108187&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=54049108187&partnerID=8YFLogxK

U2 - 10.6009/jjrt.64.915

DO - 10.6009/jjrt.64.915

M3 - Article

C2 - 18772533

AN - SCOPUS:54049108187

VL - 64

SP - 915

EP - 921

JO - Nippon Hoshasen Gijutsu Gakkai zasshi

JF - Nippon Hoshasen Gijutsu Gakkai zasshi

SN - 0369-4305

IS - 8

ER -