Abstract

PURPOSE: To determine the appropriate acquisition parameters for magnetic resonance cholangiopancreatography (MRCP) with a half-Fourier rapid acquisition with relaxation enhancement (RARE) sequence; to determine the optimal MRCP technique by comparing half-Fourier RARE, steady-state free precession (SSFP), two-dimensional (2D) fast spin-echo (SE), and three- dimensional (3D) fast SE sequences; and to clarify the usefulness and limitations of MRCP in diagnosing pancreatic abnormalities. MATERIALS AND METHODS: Half-Fourier RARE MRCP images with varying parameters were compared by using a phantom. Duct conspicuity and contrast-to-noise ratios (C/Ns) were compared for the four MRCP techniques in a phantom and healthy volunteers. The optimal MRCP technique was used to study healthy volunteers and clinical cases. Receiver operating characteristic (ROC) curves were created for data analysis. RESULTS: A 5-mm-thick section without intersection gap was appropriate for half-Fourier RARE MRCP. Only half-Fourier MRCP could depict a 1-mm duct. C/N was the highest with half-Fourier RARE, followed by 3D fast SE, 2D fast SE, and SSFP sequences. ROC curve analysis revealed no interobserver differences, and the area under the curve for detection of strictures of the main pancreatic duct was as high as 0.89. CONCLUSION: Half- Fourier RARE MRCP has the highest contrast and spatial resolution among the four techniques studied and may play an important role in diagnosing pancreatic abnormalities.

Original languageEnglish
Pages (from-to)379-387
Number of pages9
JournalRadiology
Volume206
Issue number2
DOIs
Publication statusPublished - Jan 1 1998

Fingerprint

Magnetic Resonance Cholangiopancreatography
ROC Curve
Healthy Volunteers
Pancreatic Ducts
Area Under Curve
Noise
Pathologic Constriction

All Science Journal Classification (ASJC) codes

  • Radiology Nuclear Medicine and imaging

Cite this

Irie, H., Honda, H., Tajima, T., Kuroiwa, T., Yoshimitsu, K., Makisumi, K., & Masuda, K. (1998). Optimal MR cholangiopancreatographic sequence and its clinical application. Radiology, 206(2), 379-387. https://doi.org/10.1148/radiology.206.2.9457189

Optimal MR cholangiopancreatographic sequence and its clinical application. / Irie, Hiroyuki; Honda, Hiroshi; Tajima, Tsuyoshi; Kuroiwa, Toshiro; Yoshimitsu, Kengo; Makisumi, Kenji; Masuda, Kouji.

In: Radiology, Vol. 206, No. 2, 01.01.1998, p. 379-387.

Research output: Contribution to journalArticle

Irie, H, Honda, H, Tajima, T, Kuroiwa, T, Yoshimitsu, K, Makisumi, K & Masuda, K 1998, 'Optimal MR cholangiopancreatographic sequence and its clinical application', Radiology, vol. 206, no. 2, pp. 379-387. https://doi.org/10.1148/radiology.206.2.9457189
Irie H, Honda H, Tajima T, Kuroiwa T, Yoshimitsu K, Makisumi K et al. Optimal MR cholangiopancreatographic sequence and its clinical application. Radiology. 1998 Jan 1;206(2):379-387. https://doi.org/10.1148/radiology.206.2.9457189
Irie, Hiroyuki ; Honda, Hiroshi ; Tajima, Tsuyoshi ; Kuroiwa, Toshiro ; Yoshimitsu, Kengo ; Makisumi, Kenji ; Masuda, Kouji. / Optimal MR cholangiopancreatographic sequence and its clinical application. In: Radiology. 1998 ; Vol. 206, No. 2. pp. 379-387.
@article{5aeb3cd141664b4f939b505031b43c94,
title = "Optimal MR cholangiopancreatographic sequence and its clinical application",
abstract = "PURPOSE: To determine the appropriate acquisition parameters for magnetic resonance cholangiopancreatography (MRCP) with a half-Fourier rapid acquisition with relaxation enhancement (RARE) sequence; to determine the optimal MRCP technique by comparing half-Fourier RARE, steady-state free precession (SSFP), two-dimensional (2D) fast spin-echo (SE), and three- dimensional (3D) fast SE sequences; and to clarify the usefulness and limitations of MRCP in diagnosing pancreatic abnormalities. MATERIALS AND METHODS: Half-Fourier RARE MRCP images with varying parameters were compared by using a phantom. Duct conspicuity and contrast-to-noise ratios (C/Ns) were compared for the four MRCP techniques in a phantom and healthy volunteers. The optimal MRCP technique was used to study healthy volunteers and clinical cases. Receiver operating characteristic (ROC) curves were created for data analysis. RESULTS: A 5-mm-thick section without intersection gap was appropriate for half-Fourier RARE MRCP. Only half-Fourier MRCP could depict a 1-mm duct. C/N was the highest with half-Fourier RARE, followed by 3D fast SE, 2D fast SE, and SSFP sequences. ROC curve analysis revealed no interobserver differences, and the area under the curve for detection of strictures of the main pancreatic duct was as high as 0.89. CONCLUSION: Half- Fourier RARE MRCP has the highest contrast and spatial resolution among the four techniques studied and may play an important role in diagnosing pancreatic abnormalities.",
author = "Hiroyuki Irie and Hiroshi Honda and Tsuyoshi Tajima and Toshiro Kuroiwa and Kengo Yoshimitsu and Kenji Makisumi and Kouji Masuda",
year = "1998",
month = "1",
day = "1",
doi = "10.1148/radiology.206.2.9457189",
language = "English",
volume = "206",
pages = "379--387",
journal = "Radiology",
issn = "0033-8419",
publisher = "Radiological Society of North America Inc.",
number = "2",

}

TY - JOUR

T1 - Optimal MR cholangiopancreatographic sequence and its clinical application

AU - Irie, Hiroyuki

AU - Honda, Hiroshi

AU - Tajima, Tsuyoshi

AU - Kuroiwa, Toshiro

AU - Yoshimitsu, Kengo

AU - Makisumi, Kenji

AU - Masuda, Kouji

PY - 1998/1/1

Y1 - 1998/1/1

N2 - PURPOSE: To determine the appropriate acquisition parameters for magnetic resonance cholangiopancreatography (MRCP) with a half-Fourier rapid acquisition with relaxation enhancement (RARE) sequence; to determine the optimal MRCP technique by comparing half-Fourier RARE, steady-state free precession (SSFP), two-dimensional (2D) fast spin-echo (SE), and three- dimensional (3D) fast SE sequences; and to clarify the usefulness and limitations of MRCP in diagnosing pancreatic abnormalities. MATERIALS AND METHODS: Half-Fourier RARE MRCP images with varying parameters were compared by using a phantom. Duct conspicuity and contrast-to-noise ratios (C/Ns) were compared for the four MRCP techniques in a phantom and healthy volunteers. The optimal MRCP technique was used to study healthy volunteers and clinical cases. Receiver operating characteristic (ROC) curves were created for data analysis. RESULTS: A 5-mm-thick section without intersection gap was appropriate for half-Fourier RARE MRCP. Only half-Fourier MRCP could depict a 1-mm duct. C/N was the highest with half-Fourier RARE, followed by 3D fast SE, 2D fast SE, and SSFP sequences. ROC curve analysis revealed no interobserver differences, and the area under the curve for detection of strictures of the main pancreatic duct was as high as 0.89. CONCLUSION: Half- Fourier RARE MRCP has the highest contrast and spatial resolution among the four techniques studied and may play an important role in diagnosing pancreatic abnormalities.

AB - PURPOSE: To determine the appropriate acquisition parameters for magnetic resonance cholangiopancreatography (MRCP) with a half-Fourier rapid acquisition with relaxation enhancement (RARE) sequence; to determine the optimal MRCP technique by comparing half-Fourier RARE, steady-state free precession (SSFP), two-dimensional (2D) fast spin-echo (SE), and three- dimensional (3D) fast SE sequences; and to clarify the usefulness and limitations of MRCP in diagnosing pancreatic abnormalities. MATERIALS AND METHODS: Half-Fourier RARE MRCP images with varying parameters were compared by using a phantom. Duct conspicuity and contrast-to-noise ratios (C/Ns) were compared for the four MRCP techniques in a phantom and healthy volunteers. The optimal MRCP technique was used to study healthy volunteers and clinical cases. Receiver operating characteristic (ROC) curves were created for data analysis. RESULTS: A 5-mm-thick section without intersection gap was appropriate for half-Fourier RARE MRCP. Only half-Fourier MRCP could depict a 1-mm duct. C/N was the highest with half-Fourier RARE, followed by 3D fast SE, 2D fast SE, and SSFP sequences. ROC curve analysis revealed no interobserver differences, and the area under the curve for detection of strictures of the main pancreatic duct was as high as 0.89. CONCLUSION: Half- Fourier RARE MRCP has the highest contrast and spatial resolution among the four techniques studied and may play an important role in diagnosing pancreatic abnormalities.

UR - http://www.scopus.com/inward/record.url?scp=0031952907&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0031952907&partnerID=8YFLogxK

U2 - 10.1148/radiology.206.2.9457189

DO - 10.1148/radiology.206.2.9457189

M3 - Article

C2 - 9457189

AN - SCOPUS:0031952907

VL - 206

SP - 379

EP - 387

JO - Radiology

JF - Radiology

SN - 0033-8419

IS - 2

ER -