Potentiometric Flow Injection Analysis of Glucose Using Hexacyanoferrate(III)-Hexacyanoferrate(II) Potential Buffer

Hiroki Ohura, Toshihiko Imato, Sumio Yamasaki, Nobuhiko Ishibashi

    Research output: Contribution to journalArticlepeer-review

    24 Citations (Scopus)

    Abstract

    Flow injection analysis of glucose is described, using a combination of the Potentiometrie detection with the potential buffer solution consisting of hexacyanoferrate(III)-hexacyanoferrate(II). Effects of the concentration of the potential buffer solution, reaction time and temperature are examined on sensitivity of the proposed analytical method. Glucose from 10-7 M to 10-3 M, for injections of 140 µl samples, can be determined by changing concentrations of the buffer solution from 1 × 10-5 M to 1 × 10-2 M.

    Original languageEnglish
    Pages (from-to)453-456
    Number of pages4
    Journalanalytical sciences
    Volume3
    Issue number5
    DOIs
    Publication statusPublished - Feb 1 1987

    All Science Journal Classification (ASJC) codes

    • Analytical Chemistry

    Fingerprint

    Dive into the research topics of 'Potentiometric Flow Injection Analysis of Glucose Using Hexacyanoferrate(III)-Hexacyanoferrate(II) Potential Buffer'. Together they form a unique fingerprint.

    Cite this