Potentiometric flow-injection determination of boron by using a flow-through tetrafluoroborate ion-selective poly(vinyl chloride) membrane electrode

Toshihiko Imato, Takeshi Yoshizuka, Nobuhiko Ishibashi

Research output: Contribution to journalArticle

11 Citations (Scopus)

Abstract

A highly sensitive tetrafluoroborate ion-selective poly(vinyl chloride) membrane electrode was constructed and applied to a detector in the flow-injection determination of boron. Boron at the 10 ng ml -1 level can be determined by converting it to tetrafluoroborate ion in a flow of hydrofluoric acid. A sampling rate of 30 h -1 was achieved with a relative standard deviation of less than 2%.

Original languageEnglish
Pages (from-to)139-141
Number of pages3
JournalAnalytica Chimica Acta
Volume233
Issue numberC
DOIs
Publication statusPublished - Jan 1 1990

All Science Journal Classification (ASJC) codes

  • Analytical Chemistry
  • Biochemistry
  • Environmental Chemistry
  • Spectroscopy

Fingerprint Dive into the research topics of 'Potentiometric flow-injection determination of boron by using a flow-through tetrafluoroborate ion-selective poly(vinyl chloride) membrane electrode'. Together they form a unique fingerprint.