Quantitative Perfusion Imaging with Pulsed Arterial Spin Labeling: A phantom study

Tomoyuki Noguchi, Takashi Yoshiura, Akio Hiwatashi, Osamu Togao, Koji Yamashita, Kouji Kobayashi, Futoshi Mihara, Hiroshi Honda

Research output: Contribution to journalArticle

19 Citations (Scopus)

Abstract

Purpose: Pulsed arterial spin labeling (PASL) is a magnetic resonance (MR) method for measuring cerebral blood flow. Although several validation studies for PASL in animals and humans have been reported, no reports have detailed the fundamental study of PASL using a flow phantom. We compared the true and theoretical flow rates in a flow phantom to confirm the analytical validity of quantitative perfusion imaging with Q2TIPS sequence. Methods: We built a flow phantom consisting of a 40-mm diameter plastic syringe filled with plastic beads and small plastic tubes 4 mm in diameter. Gd-DTPA-doped 8L water solution (0.1 mM) was circulated between the syringe and a tank through a plastic tube by a constant flow pump while the flow rate was adjusted between 0 and 2.61 cm/s. Q2TIPS sequence parameters were TI1 = 50 ms and TI2 = 1400 ms. Five imaging slices of 50 subtraction images were acquired sequentially in a distal-to-proximal direction using a single-shot echo planar imaging (EPI) technique. The theoretical flow rate calculated based upon the previously reported kinetic model for Q2TIPS was compared with the true flow rate. Results: A good linear relationship was observed between the theoretical, F', and true flow rates, F, in a flow rate range of 1.43 to 1.95 cm/s (F = 1.024-F- 1.915, R2 = 0.902). The ratio of theoretical to true flow rate was 92 (+/-) 4%. Conclusion: Flow rate was quantified with reasonable accuracy when the entire amount of labeled bolus within the phantom could be recovered. Our experiment confirmed the analytical validity of Q2TIPS and suggested that blood flow measurement may be feasible using the Q2TIPS pulse sequence and kinetic model of the PASL equation.

Original languageEnglish
Pages (from-to)91-97
Number of pages7
JournalMagnetic Resonance in Medical Sciences
Volume6
Issue number2
DOIs
Publication statusPublished - Dec 1 2007

Fingerprint

Perfusion Imaging
Plastics
Syringes
Cerebrovascular Circulation
Echo-Planar Imaging
Gadolinium DTPA
Validation Studies
Magnetic Resonance Spectroscopy
Water

All Science Journal Classification (ASJC) codes

  • Radiology Nuclear Medicine and imaging

Cite this

Quantitative Perfusion Imaging with Pulsed Arterial Spin Labeling : A phantom study. / Noguchi, Tomoyuki; Yoshiura, Takashi; Hiwatashi, Akio; Togao, Osamu; Yamashita, Koji; Kobayashi, Kouji; Mihara, Futoshi; Honda, Hiroshi.

In: Magnetic Resonance in Medical Sciences, Vol. 6, No. 2, 01.12.2007, p. 91-97.

Research output: Contribution to journalArticle

Noguchi, Tomoyuki ; Yoshiura, Takashi ; Hiwatashi, Akio ; Togao, Osamu ; Yamashita, Koji ; Kobayashi, Kouji ; Mihara, Futoshi ; Honda, Hiroshi. / Quantitative Perfusion Imaging with Pulsed Arterial Spin Labeling : A phantom study. In: Magnetic Resonance in Medical Sciences. 2007 ; Vol. 6, No. 2. pp. 91-97.
@article{19558e2a8d83411baeff7fd4a624d2ee,
title = "Quantitative Perfusion Imaging with Pulsed Arterial Spin Labeling: A phantom study",
abstract = "Purpose: Pulsed arterial spin labeling (PASL) is a magnetic resonance (MR) method for measuring cerebral blood flow. Although several validation studies for PASL in animals and humans have been reported, no reports have detailed the fundamental study of PASL using a flow phantom. We compared the true and theoretical flow rates in a flow phantom to confirm the analytical validity of quantitative perfusion imaging with Q2TIPS sequence. Methods: We built a flow phantom consisting of a 40-mm diameter plastic syringe filled with plastic beads and small plastic tubes 4 mm in diameter. Gd-DTPA-doped 8L water solution (0.1 mM) was circulated between the syringe and a tank through a plastic tube by a constant flow pump while the flow rate was adjusted between 0 and 2.61 cm/s. Q2TIPS sequence parameters were TI1 = 50 ms and TI2 = 1400 ms. Five imaging slices of 50 subtraction images were acquired sequentially in a distal-to-proximal direction using a single-shot echo planar imaging (EPI) technique. The theoretical flow rate calculated based upon the previously reported kinetic model for Q2TIPS was compared with the true flow rate. Results: A good linear relationship was observed between the theoretical, F', and true flow rates, F, in a flow rate range of 1.43 to 1.95 cm/s (F = 1.024-F- 1.915, R2 = 0.902). The ratio of theoretical to true flow rate was 92 (+/-) 4{\%}. Conclusion: Flow rate was quantified with reasonable accuracy when the entire amount of labeled bolus within the phantom could be recovered. Our experiment confirmed the analytical validity of Q2TIPS and suggested that blood flow measurement may be feasible using the Q2TIPS pulse sequence and kinetic model of the PASL equation.",
author = "Tomoyuki Noguchi and Takashi Yoshiura and Akio Hiwatashi and Osamu Togao and Koji Yamashita and Kouji Kobayashi and Futoshi Mihara and Hiroshi Honda",
year = "2007",
month = "12",
day = "1",
doi = "10.2463/mrms.6.91",
language = "English",
volume = "6",
pages = "91--97",
journal = "Magnetic Resonance in Medical Sciences",
issn = "1347-3182",
publisher = "Japan Society of Magnetic Resonance in Medicine",
number = "2",

}

TY - JOUR

T1 - Quantitative Perfusion Imaging with Pulsed Arterial Spin Labeling

T2 - A phantom study

AU - Noguchi, Tomoyuki

AU - Yoshiura, Takashi

AU - Hiwatashi, Akio

AU - Togao, Osamu

AU - Yamashita, Koji

AU - Kobayashi, Kouji

AU - Mihara, Futoshi

AU - Honda, Hiroshi

PY - 2007/12/1

Y1 - 2007/12/1

N2 - Purpose: Pulsed arterial spin labeling (PASL) is a magnetic resonance (MR) method for measuring cerebral blood flow. Although several validation studies for PASL in animals and humans have been reported, no reports have detailed the fundamental study of PASL using a flow phantom. We compared the true and theoretical flow rates in a flow phantom to confirm the analytical validity of quantitative perfusion imaging with Q2TIPS sequence. Methods: We built a flow phantom consisting of a 40-mm diameter plastic syringe filled with plastic beads and small plastic tubes 4 mm in diameter. Gd-DTPA-doped 8L water solution (0.1 mM) was circulated between the syringe and a tank through a plastic tube by a constant flow pump while the flow rate was adjusted between 0 and 2.61 cm/s. Q2TIPS sequence parameters were TI1 = 50 ms and TI2 = 1400 ms. Five imaging slices of 50 subtraction images were acquired sequentially in a distal-to-proximal direction using a single-shot echo planar imaging (EPI) technique. The theoretical flow rate calculated based upon the previously reported kinetic model for Q2TIPS was compared with the true flow rate. Results: A good linear relationship was observed between the theoretical, F', and true flow rates, F, in a flow rate range of 1.43 to 1.95 cm/s (F = 1.024-F- 1.915, R2 = 0.902). The ratio of theoretical to true flow rate was 92 (+/-) 4%. Conclusion: Flow rate was quantified with reasonable accuracy when the entire amount of labeled bolus within the phantom could be recovered. Our experiment confirmed the analytical validity of Q2TIPS and suggested that blood flow measurement may be feasible using the Q2TIPS pulse sequence and kinetic model of the PASL equation.

AB - Purpose: Pulsed arterial spin labeling (PASL) is a magnetic resonance (MR) method for measuring cerebral blood flow. Although several validation studies for PASL in animals and humans have been reported, no reports have detailed the fundamental study of PASL using a flow phantom. We compared the true and theoretical flow rates in a flow phantom to confirm the analytical validity of quantitative perfusion imaging with Q2TIPS sequence. Methods: We built a flow phantom consisting of a 40-mm diameter plastic syringe filled with plastic beads and small plastic tubes 4 mm in diameter. Gd-DTPA-doped 8L water solution (0.1 mM) was circulated between the syringe and a tank through a plastic tube by a constant flow pump while the flow rate was adjusted between 0 and 2.61 cm/s. Q2TIPS sequence parameters were TI1 = 50 ms and TI2 = 1400 ms. Five imaging slices of 50 subtraction images were acquired sequentially in a distal-to-proximal direction using a single-shot echo planar imaging (EPI) technique. The theoretical flow rate calculated based upon the previously reported kinetic model for Q2TIPS was compared with the true flow rate. Results: A good linear relationship was observed between the theoretical, F', and true flow rates, F, in a flow rate range of 1.43 to 1.95 cm/s (F = 1.024-F- 1.915, R2 = 0.902). The ratio of theoretical to true flow rate was 92 (+/-) 4%. Conclusion: Flow rate was quantified with reasonable accuracy when the entire amount of labeled bolus within the phantom could be recovered. Our experiment confirmed the analytical validity of Q2TIPS and suggested that blood flow measurement may be feasible using the Q2TIPS pulse sequence and kinetic model of the PASL equation.

UR - http://www.scopus.com/inward/record.url?scp=35348821877&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=35348821877&partnerID=8YFLogxK

U2 - 10.2463/mrms.6.91

DO - 10.2463/mrms.6.91

M3 - Article

C2 - 17690539

AN - SCOPUS:35348821877

VL - 6

SP - 91

EP - 97

JO - Magnetic Resonance in Medical Sciences

JF - Magnetic Resonance in Medical Sciences

SN - 1347-3182

IS - 2

ER -