Automated Estimation Method of Patient Setup Errors Using Simulated Portal Images for Prostate Cancer Radiotherapy

N. Matsushita, Hidetaka Arimura, Yoshiyuki Shioyama, T. Magome, K. Nakamura, Y. Umezu, Tadamasa Yoshitake, S. Anai, S. Yoshidome, Hiroshi Honda, Masafumi Ohki, F. Toyofuku, H. Hirata

Research output: Contribution to journalArticle

Abstract

Purpose: We developed a novel automated estimation method for patient setup errors based on simulated and real portal images for prostate cancer radiotherapy. Methods: The estimation of patient setup errors in this study was based on a template matching technique with a cross‐correlation coefficient and Sobel filter between the real portal image and localized pelvic template of reference image, which were DRR (digitally reconstructed radiography) images and simulated portal images. The simulated portal image was derived by projecting a CT image according to an inverse exponential power law of x‐ray attenuation for a water‐equivalent path length of each voxel of the CT image on each ray from a source to each pixel on the EPID (electric portal imaging device). A localized pelvic template of each patient in AP (anterior‐posterior) or lateral view was automatically extracted from the DRR or simulated portal images by cropping a rectangular region, which was determined by using the mean pelvic template and four anatomical feature points. We applied the proposed method to three prostate cancer cases, and evaluated it using the residual error between the patient setup error obtained by proposed method and the gold standard setup error determined by two radiation oncologists. Results: The average residual errors of the patient setup error for the DRR and simulated portal images were 0.79 and 1.26 mm in the left‐right (LR) direction, 3.17 and 2.05 mm in the superior‐inferior (SI) direction, 1.69 and 5.82 mm in the anterior‐posterior (AP) direction, 3.84 and 6.94 mm in Euclidean distance (ED), respectively. If we used the simulated portal image for LR and SI directions and the DRR image for AP direction, the Euclidean distance was 3.22 mm. Conclusions: The proposed method has a potential to correctly estimate patient setup errors for prostate cancer radiotherapy.

Original languageEnglish
Number of pages1
JournalMedical Physics
Volume39
Issue number6
DOIs
Publication statusPublished - Jan 1 2012

Fingerprint

Prostatic Neoplasms
Radiotherapy
Radiography
X-Rays
Direction compound
Equipment and Supplies

All Science Journal Classification (ASJC) codes

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

SU‐E‐J‐26 : Automated Estimation Method of Patient Setup Errors Using Simulated Portal Images for Prostate Cancer Radiotherapy. / Matsushita, N.; Arimura, Hidetaka; Shioyama, Yoshiyuki; Magome, T.; Nakamura, K.; Umezu, Y.; Yoshitake, Tadamasa; Anai, S.; Yoshidome, S.; Honda, Hiroshi; Ohki, Masafumi; Toyofuku, F.; Hirata, H.

In: Medical Physics, Vol. 39, No. 6, 01.01.2012.

Research output: Contribution to journalArticle

Matsushita, N. ; Arimura, Hidetaka ; Shioyama, Yoshiyuki ; Magome, T. ; Nakamura, K. ; Umezu, Y. ; Yoshitake, Tadamasa ; Anai, S. ; Yoshidome, S. ; Honda, Hiroshi ; Ohki, Masafumi ; Toyofuku, F. ; Hirata, H. / SU‐E‐J‐26 : Automated Estimation Method of Patient Setup Errors Using Simulated Portal Images for Prostate Cancer Radiotherapy. In: Medical Physics. 2012 ; Vol. 39, No. 6.
@article{d67750cd90564ae0a0cfc963fef11157,
title = "SU‐E‐J‐26: Automated Estimation Method of Patient Setup Errors Using Simulated Portal Images for Prostate Cancer Radiotherapy",
abstract = "Purpose: We developed a novel automated estimation method for patient setup errors based on simulated and real portal images for prostate cancer radiotherapy. Methods: The estimation of patient setup errors in this study was based on a template matching technique with a cross‐correlation coefficient and Sobel filter between the real portal image and localized pelvic template of reference image, which were DRR (digitally reconstructed radiography) images and simulated portal images. The simulated portal image was derived by projecting a CT image according to an inverse exponential power law of x‐ray attenuation for a water‐equivalent path length of each voxel of the CT image on each ray from a source to each pixel on the EPID (electric portal imaging device). A localized pelvic template of each patient in AP (anterior‐posterior) or lateral view was automatically extracted from the DRR or simulated portal images by cropping a rectangular region, which was determined by using the mean pelvic template and four anatomical feature points. We applied the proposed method to three prostate cancer cases, and evaluated it using the residual error between the patient setup error obtained by proposed method and the gold standard setup error determined by two radiation oncologists. Results: The average residual errors of the patient setup error for the DRR and simulated portal images were 0.79 and 1.26 mm in the left‐right (LR) direction, 3.17 and 2.05 mm in the superior‐inferior (SI) direction, 1.69 and 5.82 mm in the anterior‐posterior (AP) direction, 3.84 and 6.94 mm in Euclidean distance (ED), respectively. If we used the simulated portal image for LR and SI directions and the DRR image for AP direction, the Euclidean distance was 3.22 mm. Conclusions: The proposed method has a potential to correctly estimate patient setup errors for prostate cancer radiotherapy.",
author = "N. Matsushita and Hidetaka Arimura and Yoshiyuki Shioyama and T. Magome and K. Nakamura and Y. Umezu and Tadamasa Yoshitake and S. Anai and S. Yoshidome and Hiroshi Honda and Masafumi Ohki and F. Toyofuku and H. Hirata",
year = "2012",
month = "1",
day = "1",
doi = "10.1118/1.4734859",
language = "English",
volume = "39",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - SU‐E‐J‐26

T2 - Automated Estimation Method of Patient Setup Errors Using Simulated Portal Images for Prostate Cancer Radiotherapy

AU - Matsushita, N.

AU - Arimura, Hidetaka

AU - Shioyama, Yoshiyuki

AU - Magome, T.

AU - Nakamura, K.

AU - Umezu, Y.

AU - Yoshitake, Tadamasa

AU - Anai, S.

AU - Yoshidome, S.

AU - Honda, Hiroshi

AU - Ohki, Masafumi

AU - Toyofuku, F.

AU - Hirata, H.

PY - 2012/1/1

Y1 - 2012/1/1

N2 - Purpose: We developed a novel automated estimation method for patient setup errors based on simulated and real portal images for prostate cancer radiotherapy. Methods: The estimation of patient setup errors in this study was based on a template matching technique with a cross‐correlation coefficient and Sobel filter between the real portal image and localized pelvic template of reference image, which were DRR (digitally reconstructed radiography) images and simulated portal images. The simulated portal image was derived by projecting a CT image according to an inverse exponential power law of x‐ray attenuation for a water‐equivalent path length of each voxel of the CT image on each ray from a source to each pixel on the EPID (electric portal imaging device). A localized pelvic template of each patient in AP (anterior‐posterior) or lateral view was automatically extracted from the DRR or simulated portal images by cropping a rectangular region, which was determined by using the mean pelvic template and four anatomical feature points. We applied the proposed method to three prostate cancer cases, and evaluated it using the residual error between the patient setup error obtained by proposed method and the gold standard setup error determined by two radiation oncologists. Results: The average residual errors of the patient setup error for the DRR and simulated portal images were 0.79 and 1.26 mm in the left‐right (LR) direction, 3.17 and 2.05 mm in the superior‐inferior (SI) direction, 1.69 and 5.82 mm in the anterior‐posterior (AP) direction, 3.84 and 6.94 mm in Euclidean distance (ED), respectively. If we used the simulated portal image for LR and SI directions and the DRR image for AP direction, the Euclidean distance was 3.22 mm. Conclusions: The proposed method has a potential to correctly estimate patient setup errors for prostate cancer radiotherapy.

AB - Purpose: We developed a novel automated estimation method for patient setup errors based on simulated and real portal images for prostate cancer radiotherapy. Methods: The estimation of patient setup errors in this study was based on a template matching technique with a cross‐correlation coefficient and Sobel filter between the real portal image and localized pelvic template of reference image, which were DRR (digitally reconstructed radiography) images and simulated portal images. The simulated portal image was derived by projecting a CT image according to an inverse exponential power law of x‐ray attenuation for a water‐equivalent path length of each voxel of the CT image on each ray from a source to each pixel on the EPID (electric portal imaging device). A localized pelvic template of each patient in AP (anterior‐posterior) or lateral view was automatically extracted from the DRR or simulated portal images by cropping a rectangular region, which was determined by using the mean pelvic template and four anatomical feature points. We applied the proposed method to three prostate cancer cases, and evaluated it using the residual error between the patient setup error obtained by proposed method and the gold standard setup error determined by two radiation oncologists. Results: The average residual errors of the patient setup error for the DRR and simulated portal images were 0.79 and 1.26 mm in the left‐right (LR) direction, 3.17 and 2.05 mm in the superior‐inferior (SI) direction, 1.69 and 5.82 mm in the anterior‐posterior (AP) direction, 3.84 and 6.94 mm in Euclidean distance (ED), respectively. If we used the simulated portal image for LR and SI directions and the DRR image for AP direction, the Euclidean distance was 3.22 mm. Conclusions: The proposed method has a potential to correctly estimate patient setup errors for prostate cancer radiotherapy.

UR - http://www.scopus.com/inward/record.url?scp=85024827033&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85024827033&partnerID=8YFLogxK

U2 - 10.1118/1.4734859

DO - 10.1118/1.4734859

M3 - Article

VL - 39

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -