Tetraphenylborate ion selective electrode based on plasticized poly(vinyl chloride) membrane containing no ion-exchange site and its application to potentiometric titration

Takashi Masadome, Toshihiko Imato, Nobuhiko Ishibasm

    Research output: Contribution to journalArticlepeer-review

    11 Citations (Scopus)

    Abstract

    The o-nitrophenyl octyl ether (o-NPOE)-plasticized PVC membrane electrode shows the linear response with potential slope of 50 mV per decade for tetraphenylborate (TPB-) ion from 10-5 M to 10-2 M. The response time of the electrode is in the order of a few seconds. The electrode can be used in wide pH range, and is highly selective to the TPB- ion over inorganic anions. The electrode is suitable as the end-point detector in the potentiometric titration of quaternary ammonium salts and potassium ion.

    Original languageEnglish
    Pages (from-to)508-510
    Number of pages3
    Journalbunseki kagaku
    Volume36
    Issue number8
    DOIs
    Publication statusPublished - 1987

    All Science Journal Classification (ASJC) codes

    • Analytical Chemistry

    Fingerprint

    Dive into the research topics of 'Tetraphenylborate ion selective electrode based on plasticized poly(vinyl chloride) membrane containing no ion-exchange site and its application to potentiometric titration'. Together they form a unique fingerprint.