Ultraviolet and Deep-Ultraviolet Excitation Photothermal Heterodyne Interferometer Combined with Semi-Micro HPLC

Miki Isoda, Toshihiko Abe, Kai Aoki, Akira Harata

Research output: Contribution to journalArticlepeer-review

Abstract

A detection system consisting of a photothermal heterodyne interferometer (PHI) combined with semi-micro HPLC (high-performance liquid chromatography) has been designed and investigated. An ultraviolet (UV) or deep-UV laser emitting at 375 or 213 nm, respectively, was used for the excitation of nitro-polycyclic aromatic hydrocarbons (NPAHs) and amino acids. A photothermally induced change in the refractive index of the solvent causes an optical phase difference between two arms of the interferometer, one beam passing through the photoexcited region and another used as a reference, which was sensitively detected with the PHI. The separation and detection of NPAHs and amino acids were successfully demonstrated via semi-micro HPLC with the PHI and a UV detector. The detection limits of the UVexcitation PHI for NPAHs were 1.2 – 5.2 times better than that of the commercial UV detector, although the first demonstration of deep-UV excitation suffered from significant baseline fluctuation.

Original languageEnglish
Pages (from-to)911-916
Number of pages6
Journalanalytical sciences
Volume37
Issue number6
DOIs
Publication statusPublished - 2021

All Science Journal Classification (ASJC) codes

  • Analytical Chemistry

Fingerprint

Dive into the research topics of 'Ultraviolet and Deep-Ultraviolet Excitation Photothermal Heterodyne Interferometer Combined with Semi-Micro HPLC'. Together they form a unique fingerprint.