T1 - Use of an inertial measurement unit sensor in pedicle screw placement improves trajectory accuracy
AU - Baba, Satoshi
AU - Kawaguchi, Kenichi
AU - Itamoto, Kazuhito
AU - Watanabe, Takeshi
AU - Hayashida, Mitsumasa
AU - Mae, Takao
AU - Nakashima, Yasuharu
AU - Kato, Go
N1 - Funding Information:
This work was supported by ZENKYOREN (National Mutual Insurance of Federation of Agricultural Cooperatives)(https:// social.ja-kyosai.or.jp/contribution/traffic06.html) to G.K. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript. The authors would like to thank Enago (www.enago.jp) for English language review.
Publisher Copyright:
© 2020 Baba et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
PY - 2020/11
Y1 - 2020/11
N2 - Ascertaining the accuracy of the pedicle screw (PS) trajectories is important as PS malpositioning can cause critical complications. We aimed to determine the angle range over which estimation is unreliable; build a low-cost PS placement support system that uses an inertial measurement unit (IMU) to enable the monitoring of surgical tools and PS trajectories, and determine the situations where IMU support would be most beneficial. In PS insertion experiments, we used cadaver samples that included lumbar porcine spines. Computed tomography images obtained before and after PS insertion were viewed. Offsets between the planned and implanted PS trajectories in the freehand and IMU-assisted groups were analyzed. The PS cortical bone breaches were classified according to the Gertzbein and Robbins criteria (GRC). Added head-down tilted sample experiments were repeated wherein we expected a decreased rostro-caudal rotational accuracy of the PS according to the angle estimation ability results. Evaluation of the PS trajectory accuracy revealed no significant advantage of IMU-assisted rostro-caudal rotational accuracy versus freehand accuracy. According to the GRC, IMU assistance significantly increased the rate of clinically acceptable PS positions (RoCA) than the freehand technique. In the head-down tilted sample experiments, IMU assist provided increased accuracies with both rostro-caudal and medial rotational techniques when compared with the freehand technique. In the freehand group, RoCA was significantly decreased in samples with rostral tilting relative to that in the samples without. However, In the IMU-assisted group, no significant difference in RoCA between the samples with and without head-down tilting was observed. Even when the planned PS medial and/or rostro-caudal rotational angle was relatively large and difficult to reproduce manually, IMU-support helped maintain the PS trajectory accuracy and positioning safety. IMU assist in PS placement was more beneficial, especially for larger rostro-caudal and/or medial rotational pedicle angles.
AB - Ascertaining the accuracy of the pedicle screw (PS) trajectories is important as PS malpositioning can cause critical complications. We aimed to determine the angle range over which estimation is unreliable; build a low-cost PS placement support system that uses an inertial measurement unit (IMU) to enable the monitoring of surgical tools and PS trajectories, and determine the situations where IMU support would be most beneficial. In PS insertion experiments, we used cadaver samples that included lumbar porcine spines. Computed tomography images obtained before and after PS insertion were viewed. Offsets between the planned and implanted PS trajectories in the freehand and IMU-assisted groups were analyzed. The PS cortical bone breaches were classified according to the Gertzbein and Robbins criteria (GRC). Added head-down tilted sample experiments were repeated wherein we expected a decreased rostro-caudal rotational accuracy of the PS according to the angle estimation ability results. Evaluation of the PS trajectory accuracy revealed no significant advantage of IMU-assisted rostro-caudal rotational accuracy versus freehand accuracy. According to the GRC, IMU assistance significantly increased the rate of clinically acceptable PS positions (RoCA) than the freehand technique. In the head-down tilted sample experiments, IMU assist provided increased accuracies with both rostro-caudal and medial rotational techniques when compared with the freehand technique. In the freehand group, RoCA was significantly decreased in samples with rostral tilting relative to that in the samples without. However, In the IMU-assisted group, no significant difference in RoCA between the samples with and without head-down tilting was observed. Even when the planned PS medial and/or rostro-caudal rotational angle was relatively large and difficult to reproduce manually, IMU-support helped maintain the PS trajectory accuracy and positioning safety. IMU assist in PS placement was more beneficial, especially for larger rostro-caudal and/or medial rotational pedicle angles.
UR - http://www.scopus.com/inward/record.url?scp=85096249571&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85096249571&partnerID=8YFLogxK
U2 - 10.1371/journal.pone.0242512
DO - 10.1371/journal.pone.0242512
M3 - Article
C2 - 33196657
AN - SCOPUS:85096249571
SN - 1932-6203
VL - 15
JO - PLoS One
JF - PLoS One
IS - 11 November
M1 - e0242512
ER -