A thin plastic scintillator is used as a trigger counter in a charged particle spectrometer. Since its Compton edge is not clear, it is difficult to test the thin plastic scintillator counter with a standard gamma-ray source. In this paper, a simple method is proposed to determine the Compton edge and the energy resolution of the thin plastic scintillator by fitting the simulated spectrum to the measured spectrum. The simulated spectrum was made with a Monte Carlo program utilizing the geometry of the plastic scintillator. This technique was examined with three different plastic scintillators.

元の言語英語
ページ(範囲)1172-1176
ページ数5
ジャーナルIEEE Transactions on Nuclear Science
48
発行部数4 I
DOI
出版物ステータス出版済み - 8 1 2001
イベントNuclear Science Symposium (NSS) 2000 - Lyon, フランス
継続期間: 10 15 200010 20 2000

Fingerprint

Phosphors
scintillation counters
plastics
Calibration
Plastics
counters
Particle spectrometers
energy
Charged particles
Gamma rays
charged particles
actuators
spectrometers
Geometry
geometry

All Science Journal Classification (ASJC) codes

  • Nuclear and High Energy Physics
  • Nuclear Energy and Engineering
  • Electrical and Electronic Engineering

これを引用

A simple method of energy calibration for thin plastic scintillator. / Hohara, S.; Saiho, F.; Tanaka, J.; Aoki, S.; Uozumi, Y.; Matoba, M.

:: IEEE Transactions on Nuclear Science, 巻 48, 番号 4 I, 01.08.2001, p. 1172-1176.

研究成果: ジャーナルへの寄稿Conference article

Hohara, S, Saiho, F, Tanaka, J, Aoki, S, Uozumi, Y & Matoba, M 2001, 'A simple method of energy calibration for thin plastic scintillator', IEEE Transactions on Nuclear Science, 巻. 48, 番号 4 I, pp. 1172-1176. https://doi.org/10.1109/23.958745
Hohara, S. ; Saiho, F. ; Tanaka, J. ; Aoki, S. ; Uozumi, Y. ; Matoba, M. / A simple method of energy calibration for thin plastic scintillator. :: IEEE Transactions on Nuclear Science. 2001 ; 巻 48, 番号 4 I. pp. 1172-1176.
@article{409fdf5f66864b52bbb5767d55a86a86,
title = "A simple method of energy calibration for thin plastic scintillator",
abstract = "A thin plastic scintillator is used as a trigger counter in a charged particle spectrometer. Since its Compton edge is not clear, it is difficult to test the thin plastic scintillator counter with a standard gamma-ray source. In this paper, a simple method is proposed to determine the Compton edge and the energy resolution of the thin plastic scintillator by fitting the simulated spectrum to the measured spectrum. The simulated spectrum was made with a Monte Carlo program utilizing the geometry of the plastic scintillator. This technique was examined with three different plastic scintillators.",
author = "S. Hohara and F. Saiho and J. Tanaka and S. Aoki and Y. Uozumi and M. Matoba",
year = "2001",
month = "8",
day = "1",
doi = "10.1109/23.958745",
language = "English",
volume = "48",
pages = "1172--1176",
journal = "IEEE Transactions on Nuclear Science",
issn = "0018-9499",
publisher = "Institute of Electrical and Electronics Engineers Inc.",
number = "4 I",

}

TY - JOUR

T1 - A simple method of energy calibration for thin plastic scintillator

AU - Hohara, S.

AU - Saiho, F.

AU - Tanaka, J.

AU - Aoki, S.

AU - Uozumi, Y.

AU - Matoba, M.

PY - 2001/8/1

Y1 - 2001/8/1

N2 - A thin plastic scintillator is used as a trigger counter in a charged particle spectrometer. Since its Compton edge is not clear, it is difficult to test the thin plastic scintillator counter with a standard gamma-ray source. In this paper, a simple method is proposed to determine the Compton edge and the energy resolution of the thin plastic scintillator by fitting the simulated spectrum to the measured spectrum. The simulated spectrum was made with a Monte Carlo program utilizing the geometry of the plastic scintillator. This technique was examined with three different plastic scintillators.

AB - A thin plastic scintillator is used as a trigger counter in a charged particle spectrometer. Since its Compton edge is not clear, it is difficult to test the thin plastic scintillator counter with a standard gamma-ray source. In this paper, a simple method is proposed to determine the Compton edge and the energy resolution of the thin plastic scintillator by fitting the simulated spectrum to the measured spectrum. The simulated spectrum was made with a Monte Carlo program utilizing the geometry of the plastic scintillator. This technique was examined with three different plastic scintillators.

UR - http://www.scopus.com/inward/record.url?scp=0035428747&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0035428747&partnerID=8YFLogxK

U2 - 10.1109/23.958745

DO - 10.1109/23.958745

M3 - Conference article

AN - SCOPUS:0035428747

VL - 48

SP - 1172

EP - 1176

JO - IEEE Transactions on Nuclear Science

JF - IEEE Transactions on Nuclear Science

SN - 0018-9499

IS - 4 I

ER -