A theoretical model to predict tensile deformation behavior of balloon catheter

Mitsugu Todo, Keiji Yoshiya, Takuya Matsumoto

研究成果: ジャーナルへの寄稿学術誌査読

抄録

In this technical note, a simple theoretical model was proposed to express the tensile deformation and fracture of balloon catheter tested by the ISO standard using piece-wise linear force-displacement relations. The model was then validated by comparing with the tensile force-displacement behaviors of two types of typical balloon catheters clinically used worldwide. It was shown that the proposed model can effectively be used to express the tensile deformation behavior and easily be handled by physicians who are not familiar with mechanics of materials.

本文言語英語
ページ(範囲)236-239
ページ数4
ジャーナルJournal of the Mechanical Behavior of Biomedical Materials
62
DOI
出版ステータス出版済み - 9月 1 2016