An evaluation of computed tomography dose index measurements using a pencil ionisation chamber and small detectors

Choirul Anam, Toshioh Fujibuchi, Freddy Haryanto, Rena Widita, Idam Arif, Geoff Dougherty

研究成果: Contribution to journalArticle査読

9 被引用数 (Scopus)

抄録

The aim of this study was to compare the values of the computed tomography dose index 100 (CTDI100) obtained using two small detectors (i.e. a small ionisation chamber and a small solid state detector) with those obtained from a 100 mm pencil ionisation chamber for various input CT parameters: beam width, kVp, mAs, pitch, and head-body phantom variation. The measurement of CTDI100 using the 100 mm pencil chamber was carried out in a single rotation of axial mode, while the measurement using small detectors was carried out in helical mode. The differences of CTDI100 values obtained with two small detectors were about 7% for all variations. The differences of CTDI100 values obtained with small detectors and a 100 mm pencil ionisation chamber for beam widths of more than 4 mm were within 40%. However, for the narrowest beam widths (4 mm), the difference between them was very large (about 150%).

本文言語英語
ページ(範囲)112-124
ページ数13
ジャーナルJournal of Radiological Protection
39
1
DOI
出版ステータス出版済み - 3 2019

All Science Journal Classification (ASJC) codes

  • 廃棄物管理と処理
  • 公衆衛生学、環境および労働衛生

フィンガープリント

「An evaluation of computed tomography dose index measurements using a pencil ionisation chamber and small detectors」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。