An evaluation of the basic characteristics of a plastic scintillating fibre detector in CT radiation fields

Kento Terasaki, Toshioh Fujibuchi, Takatoshi Toyoda, Yutaka Yoshida, Tsutomu Akasaka, Akihiro Nohtomi, Junji Morishita

研究成果: Contribution to journalArticle査読

1 被引用数 (Scopus)

抄録

The ionisation chamber for computed tomography (CT) is an instrument that is most commonly used to measure the computed tomography dose index. However, it has been reported that the 10 cm effective detection length of the ionisation chamber is insufficient due to the extent of the dose distribution outside the chamber. The purpose of this study was to estimate the basic characteristics of a plastic scintillating fibre (PSF) detector with a long detection length of 50 cm in CT radiation fields. The authors investigated position dependence using diagnostic X-ray equipment and dependencies for energy, dose rate and slice thickness using an X-ray CT system. The PSF detector outputs piled up at a count rate of 10 000 counts ms-1 in dose rate dependence study. With calibration, this detector may be useful as a CT dosemeter with a long detection length except for the measurement at high dose rate.

本文言語英語
ページ(範囲)527-533
ページ数7
ジャーナルRadiation Protection Dosimetry
171
4
DOI
出版ステータス出版済み - 2016