Anion Selective Electrode for High Performance Liquid Chromatography Detector Based on Oleophilic Anion-Exchange Resin Membrane

Sadayuki Koizumi, Toshihiko Imato, Nobuhiko Ishibashi

研究成果: Contribution to journalArticle査読

8 被引用数 (Scopus)

抄録

A system is described for the separate determination of anions using HPLC with a potentiometric detector based on an ion selective electrode (ISE). The ISE consists of an oleophilic anion-exchange resin membrane impregnated with an organic solvent. A sodium sulfate solution containing a trace level of a primary ion for the electrode was used as an eluent. Sensitivity and selectivity of the ISE detector varies with a solvent species imbibed in the detector membrane and an ionic form of the membrane. Chromatograms for oxyacid anions (IO3-, BrO3-, CIO3- and NO3-) are shown, and relationships between chromatographic sensitivity and selectivity coefficients of the ISE detector are discussed.

本文言語英語
ページ(範囲)319-325
ページ数7
ジャーナルanalytical sciences
3
4
DOI
出版ステータス出版済み - 1987