[Construction and usefulness of the dose management system using DICOM radiation dose structured report (Dose SR) in angiography]

Tokitaka Ueno, Yuki Emoto, Junji Morishita, Masashi Shimizu, Ryotaro Jingu, Takaaki Imasato, Shigeki Ono, Satoshi Nakashima, Hironori Nakamura, Tadahisa Uemura, Kengo Yoshimitsu

研究成果: Contribution to journalArticle査読

抄録

The International Commission on Radiological Protection recommends diagnostic reference levels (DRL) in each radiological examination for justification and optimization of patients' dose in medicine. The aim of our study was to propose the dose management system by utilizing dose information in diagnostic X-ray radiation dose structured report (Dose SR) in The Digital Imaging and Communications in Medicine to optimize radiation dose in institutions. Our dose management system is able to organize dose information obtained from various angiography systems and CTs. It is possible to provide this information to operators for justification and optimization of patient dose. Our system would be useful for the estimation of organ dose and could be used for the determination of local DRL (LDRL) for each radiological practice. In addition, the optimization became possible to compare LDRL with national DRL.

本文言語英語
ページ(範囲)1392-1402
ページ数11
ジャーナルNihon Hoshasen Gijutsu Gakkai zasshi
70
12
DOI
出版ステータス出版済み - 12 1 2014

All Science Journal Classification (ASJC) codes

  • 医学(全般)

フィンガープリント

「[Construction and usefulness of the dose management system using DICOM radiation dose structured report (Dose SR) in angiography]」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル