Design of pickup coil made of litz wire and cooled at 77K for high sensitive measurement of AC magnetic fields

Keiji Enpuku, Shintaro Hirakawa, Ryuji Momotomi, Masaaki Matsuo, Takashi Yoshida, Akihiko Kandori

研究成果: Contribution to journalArticle査読

3 被引用数 (Scopus)

抄録

We studied the performance of a copper pickup coil for high sensitive measurements of ac magnetic fields. The coil was made of Litz wire to revent the skin effect, and was cooled at 77 K to reduce thermal noise. We first obtained an analytical expression for the coil resistance by nalyzing the eddy current loss of the coil. The analytical result explains well the measured dependences of the coil resistance on the frequency, umber of turns of the coil (N), and filament diameter of the Litz wire (df). Next, we show the design of the pickup coil. When appropriate values re selected for N and df, we can expect the magnetic field noise SB 1/2 > 10 fT/Hz1/2 for f > 1 kHz and SB =2 > 2 fT/Hz1/2 for f < 10 kHz for the ickup coil with diameter of 50 mm. The experimental results agreed with the designed values.

本文言語英語
論文番号076602
ジャーナルJapanese journal of applied physics
50
7 PART 1
DOI
出版ステータス出版済み - 7 2011

All Science Journal Classification (ASJC) codes

  • 工学(全般)
  • 物理学および天文学(全般)

フィンガープリント

「Design of pickup coil made of litz wire and cooled at 77K for high sensitive measurement of AC magnetic fields」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。