Development of a novel straining holder for transmission electron microscopy compatible with single tilt-axis electron tomography

K. Sato, H. Miyazaki, T. Gondo, S. Miyazaki, M. Murayama, S. Hata

研究成果: Contribution to journalArticle査読

9 被引用数 (Scopus)

抄録

We have developed a newly designed straining specimen holder for in situ transmission electron microscopy (TEM) compatible with high-angle single tilt-axis electron tomography. The holder can deform a TEM specimen under tensile stress with the strain rate between 1.5 × 10−6 and 5.2 × 10−3 s−1. We have also confirmed that the maximum tilt angle of the specimen holder reaches ±60° with a rectangular shape aluminum specimen. The new specimen holder, termed as ‘straining and tomography holder’, will have wide range potential applications in materials science.