Effect of the concentrations of triphenyltin chloride and anion exchanger on the anion selectivity of a plasticized PVC membrane electrode

Sadayuki Koizumi, Toshihiko Imato, Nobuhiko Ishibashi

研究成果: Contribution to journalArticle

7 被引用数 (Scopus)

抄録

Potentiometric selectivities of a plasticized poly(vinyl chloride) (PVC) membrane containing a hydrophobic quaternary ammonium salt and triphenyltin chloride (TPTCl) for several inorganic anions were evaluated by measurements of the membrane potential of a bi-ionic system (the so-called bi-ionic membrane potential). The addition of TPTCl to the plasticized PVC membrane containing a quaternary ammonium salt gave rise to selectivity patterns different from the Hofmeister anion series, which is observed for a membrane containing only the quaternary ammonium salt. The change in the selectivity pattern from the Hofmeister to the non-Hofmeister was observed under conditions where the TPTCl concentration was higher than the quaternary ammonium salt in the membrane. Response mechanisms and the selectivity of the TPTCl-based electrode are discussed with respect to the concentration dependence of TPTCl and quaternary ammonium salt in the membrane.

本文言語英語
ページ(範囲)37-43
ページ数7
ジャーナルFresenius' Journal of Analytical Chemistry
357
1
DOI
出版ステータス出版済み - 1 1 1997

All Science Journal Classification (ASJC) codes

  • Biochemistry

フィンガープリント 「Effect of the concentrations of triphenyltin chloride and anion exchanger on the anion selectivity of a plasticized PVC membrane electrode」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。