Extremely low frequency magnetic fields originating from equipment used for assisted reproduction, umbilical cord and peripheral blood stem cell transplantation, transfusion, and hemodialysis

Kosei Moriyama, Hiroyuki Sato, Kenji Tanaka, Yoshiyuki Nakashima, Kuniaki Yoshitomi

研究成果: Contribution to journalArticle

5 引用 (Scopus)

抜粋

The extremely low-frequency (ELF) magnetic fields (MFs) originating from equipment used for assisted reproduction, umbilical cord-blood and peripheral-blood stem cell transplantation, transfusion, and hemodialysis were measured. The ELF-MF values were 0.1-1.2 μT on clean benches, <0.1-8.0 μT inverted microscopes, <0.1-13.6 μT in CO2 incubators, 4.3-11.5 μT in centrifuges, 0.4-18.8 μT programmed freezers, <0.1-0.3 μT deep freezers, 0.3-3.1 μT cell separators, and 0.2-0.9 μT in hemodialysers. Frequencies of MFs were nominally 60 Hz, but some devices showed non-sinusoidal 120 Hz. Such MFs can be reduced by shielding the sources or altering the protocols employed.

元の言語英語
ページ(範囲)69-73
ページ数5
ジャーナルBioelectromagnetics
26
発行部数1
DOI
出版物ステータス出版済み - 1 1 2005

All Science Journal Classification (ASJC) codes

  • Biophysics
  • Physiology
  • Radiology Nuclear Medicine and imaging

フィンガープリント Extremely low frequency magnetic fields originating from equipment used for assisted reproduction, umbilical cord and peripheral blood stem cell transplantation, transfusion, and hemodialysis' の研究トピックを掘り下げます。これらはともに一意のフィンガープリントを構成します。