Fast tomography using quasi-monochromatic undulator radiation

Kentaro Uesugi, Toshihiro Sera, Naoto Yagi

研究成果: Contribution to journalArticle査読

26 被引用数 (Scopus)

抄録

A beamline with a helical undulator has been used without a monochromator for fast high-resolution tomographic imaging with an X-ray energy of 12.4-16.5 keVand an energy bandwidth of 2-3%. The X-ray beam was expanded with two mirrors to 12 mm × 4 mm. The X-ray field was made uniform by a diffuser. The detector pixel size was 9.9 μm × 9.9 μm. At the highest speed, a 180° scan was completed in 6 s with 454 projections. Beam-hardening effects were not significant. This technique may be useful in studying time-dependent structural changes of soft materials such as polymers and biological samples.

本文言語英語
ページ(範囲)403-407
ページ数5
ジャーナルJournal of Synchrotron Radiation
13
5
DOI
出版ステータス出版済み - 9 1 2006
外部発表はい

All Science Journal Classification (ASJC) codes

  • Radiation
  • Nuclear and High Energy Physics
  • Instrumentation