抄録
A reliable and sensitive method has been developed to assess the concentrations of the hypnotic drug triazolam in human tissues, including putrefied tissues. The method involves a 3-step solvent extraction, cleanup on a silica gel column and gas chromatography using a nitrogen phosphorus detector and a capillary column. Estazolam was used as an internal standard. The calibration curve was linear over the concentration range 1 ng/g1 gm/g and the lower limit of detection was 0.5 ng/g. A forensic study was performed on the toxicological effects of triazolam using putrefied tissues.
本文言語 | 英語 |
---|---|
ページ(範囲) | 67-69 |
ページ数 | 3 |
ジャーナル | International Journal of Legal Medicine |
巻 | 104 |
号 | 2 |
DOI | |
出版ステータス | 出版済み - 3月 1 1991 |
!!!All Science Journal Classification (ASJC) codes