Improvement of in situ glass electrode pH sensor for seawater by rapid response glass electrode and removable Ag/AgCl electrode with solid reference junction using vycol glass

Kei Okamura, Takuroh Noguchi, Mayumi Hatta, Jun Ichiro Ishibashi

研究成果: Chapter in Book/Report/Conference proceedingConference contribution

抜粋

The glass membrane pH sensor PH-09 using commercial available glass membrane with separated reference electrode having pressure range up to 3400 m in depth was developed. PH-09 showed good repeatability up to 3400m. Comparison to other commercial glass membrane pH sensor showed good relationship and faster response. Comparison to colorimetric analysis showed lower pH value around -0.15.

元の言語英語
ホスト出版物のタイトルTechno-Ocean 2016
ホスト出版物のサブタイトルReturn to the Oceans
出版者Institute of Electrical and Electronics Engineers Inc.
ページ410-414
ページ数5
ISBN(電子版)9781509024452
DOI
出版物ステータス出版済み - 3 30 2017
イベント16th Techno-Ocean, Techno-Ocean 2016 - Kobe, 日本
継続期間: 10 6 201610 8 2016

出版物シリーズ

名前Techno-Ocean 2016: Return to the Oceans

その他

その他16th Techno-Ocean, Techno-Ocean 2016
日本
Kobe
期間10/6/1610/8/16

All Science Journal Classification (ASJC) codes

  • Oceanography
  • Water Science and Technology
  • Energy Engineering and Power Technology
  • Ocean Engineering
  • Instrumentation

フィンガープリント Improvement of in situ glass electrode pH sensor for seawater by rapid response glass electrode and removable Ag/AgCl electrode with solid reference junction using vycol glass' の研究トピックを掘り下げます。これらはともに一意のフィンガープリントを構成します。

  • これを引用

    Okamura, K., Noguchi, T., Hatta, M., & Ishibashi, J. I. (2017). Improvement of in situ glass electrode pH sensor for seawater by rapid response glass electrode and removable Ag/AgCl electrode with solid reference junction using vycol glass. : Techno-Ocean 2016: Return to the Oceans (pp. 410-414). [7890688] (Techno-Ocean 2016: Return to the Oceans). Institute of Electrical and Electronics Engineers Inc.. https://doi.org/10.1109/Techno-Ocean.2016.7890688