T1 - Left atrial assist device to treat patients with heart failure with preserved ejection fraction
T2 - Initial in vitro study
AU - Fukamachi, Kiyotaka
AU - Horvath, David J.
AU - Karimov, Jamshid H.
AU - Kado, Yuichiro
AU - Miyamoto, Takuma
AU - Kuban, Barry D.
AU - Starling, Randall C.
N1 - Funding Information:
This study was supported by Cleveland Clinic internal funding from the Department of Biomedical Engineering, Lerner Research Institute.
Publisher Copyright:
© 2020 The American Association for Thoracic Surgery
PY - 2021/7
Y1 - 2021/7
N2 - Objectives: Many patients with heart failure have preserved ejection fraction but also diastolic dysfunction, with no effective therapy. We are developing a new pump (left atrial assist device, LAAD) for implantation at the mitral position to pump blood from the left atrium to sufficiently fill the left ventricle. The purpose of the initial in vitro study was to demonstrate that the LAAD can reduce left atrial pressure (LAP) and increase cardiac output (CO) while maintaining arterial pulsatility and normal aortic valve function using a proof-of-concept device. Methods: The LAAD concept was tested at 3 pump speeds on a pulsatile mock loop with a pneumatic pump that simulated the normal function of the native ventricle as well as 3 levels of diastolic heart failure (DHF 1, 2, and 3) by adjusting the diastolic drive pressure to limit diastolic filling of the ventricle. Results: Without the LAAD, CO and aortic pressure (AoP) decreased dramatically from 3.8 L/min and 100 mm Hg at normal heart condition to 1.2 L/min and 35 mm Hg at DHF 3, respectively. With LAAD support, both CO and AoP recovered to normal heart values at 3200 rpm and surpassed normal heart values at 3800 rpm. Furthermore, with LAAD support, LAP recovered to almost that of the normal heart condition at 3800 rpm. Conclusions: These initial in vitro results support our hypothesis that use of the LAAD increases CO and AoP and decreases LAP under DHF conditions while maintaining arterial pulsatility and full function of the aortic valve.
AB - Objectives: Many patients with heart failure have preserved ejection fraction but also diastolic dysfunction, with no effective therapy. We are developing a new pump (left atrial assist device, LAAD) for implantation at the mitral position to pump blood from the left atrium to sufficiently fill the left ventricle. The purpose of the initial in vitro study was to demonstrate that the LAAD can reduce left atrial pressure (LAP) and increase cardiac output (CO) while maintaining arterial pulsatility and normal aortic valve function using a proof-of-concept device. Methods: The LAAD concept was tested at 3 pump speeds on a pulsatile mock loop with a pneumatic pump that simulated the normal function of the native ventricle as well as 3 levels of diastolic heart failure (DHF 1, 2, and 3) by adjusting the diastolic drive pressure to limit diastolic filling of the ventricle. Results: Without the LAAD, CO and aortic pressure (AoP) decreased dramatically from 3.8 L/min and 100 mm Hg at normal heart condition to 1.2 L/min and 35 mm Hg at DHF 3, respectively. With LAAD support, both CO and AoP recovered to normal heart values at 3200 rpm and surpassed normal heart values at 3800 rpm. Furthermore, with LAAD support, LAP recovered to almost that of the normal heart condition at 3800 rpm. Conclusions: These initial in vitro results support our hypothesis that use of the LAAD increases CO and AoP and decreases LAP under DHF conditions while maintaining arterial pulsatility and full function of the aortic valve.
UR - http://www.scopus.com/inward/record.url?scp=85079834984&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85079834984&partnerID=8YFLogxK
U2 - 10.1016/j.jtcvs.2019.12.110
DO - 10.1016/j.jtcvs.2019.12.110
M3 - Article
C2 - 32089348
AN - SCOPUS:85079834984
VL - 162
SP - 120
EP - 126
JO - Journal of Thoracic and Cardiovascular Surgery
JF - Journal of Thoracic and Cardiovascular Surgery
SN - 0022-5223
IS - 1
ER -