Lung-to-finger circulation time can be measured stably with high reproducibility by simple breath holding method in cardiac patients

Tomoyuki Tobushi, Takatoshi Kasai, Masayuki Hirose, Kazuhiro Sakai, Manabu Akamatsu, Chizuru Ohsawa, Yasuko Yoshioka, Shoko Suda, Nanako Shiroshita, Ryo Nakamura, Toshiaki Kadokami, Takeshi Tohyama, Kouta Funakoshi, Kazuya Hosokawa, Shin ichi Ando

研究成果: Contribution to journalArticle査読

抄録

Lung to finger circulation time (LFCT) has been used to estimate cardiac function. We developed a new LFCT measurement device using a laser sensor at fingertip. We measured LFCT by measuring time from re-breathing after 20 s of breath hold to the nadir of the difference of transmitted red light and infrared light, which corresponds to percutaneous oxygen saturation. Fifty patients with heart failure were enrolled. The intrasubject stability of the measurement was assessed by the intraclass correlation coefficient (ICC). The ICC calculated from 44 cases was 0.85 (95% confidence interval: 0.77–0.91), which means to have “Excellent reliability.” By measuring twice, at least one clear LFCT value was obtained in 89.1% of patients and the overall measurability was 95.7%. We conducted all LFCT measurements safely. High ICCs were obtained even after dividing patients according to age, cardiac index (CI); 0.85 and 0.84 (≥ 75 or < 75 years group, respectively), 0.81 and 0.84 (N = 26, ≥ or < 2.2 L/min/M2). These results show that our new method to measure LFCT is highly stable and feasible for any type of heart failure patients.

本文言語英語
論文番号15913
ジャーナルScientific reports
11
1
DOI
出版ステータス出版済み - 12 2021

All Science Journal Classification (ASJC) codes

  • 一般

フィンガープリント

「Lung-to-finger circulation time can be measured stably with high reproducibility by simple breath holding method in cardiac patients」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル