Measurement of activity distribution using photostimulable phosphor imaging plates in decommissioned 10 mv medical linear accelerator

Toshioh Fujibuchi, Shunsuke Yonai, Masahiro Yoshida, Takeji Sakae, Hiroshi Watanabe, Yoshihisa Abe, Jun Itami

研究成果: ジャーナルへの寄稿学術誌査読

3 被引用数 (Scopus)

抄録

Photonuclear reactions generate neutrons in the head of the linear accelerator. Therefore, some parts of the linear accelerator can become activated. Such activated materials must be handled as radioactive waste. The authors attempted to investigate the distribution of induced radioactivity using photostimulable phosphor imaging plates. Autoradiographs were produced from some parts of the linear accelerator (the target, upper jaw, multileaf collimator and shielding). The levels of induced radioactivity were confirmed to be non-uniform within each part from the autoradiographs. The method was a simple and highly sensitive approach to evaluating the relative degree of activation of the linear accelerators, so that appropriate materials management procedures can be carried out.

本文言語英語
ページ(範囲)S158-S162
ジャーナルHealth physics
107
SUPPL. 2
DOI
出版ステータス出版済み - 8月 2014