Measurement of basic characteristics of scintillation-type radiation survey meters with multi-pixel photon counter

Toshioh Fujibuchi, Takatoshi Toyoda, Kento Terasaki

研究成果: ジャーナルへの寄稿記事

1 引用 (Scopus)

抄録

The basic characteristics of three low-cost radiation survey meters using multi-pixel photon counters (MPPC), the C12137, T-GMK2-S, and iMetry, were measured. The linearity of the dose rate was confirmed over the full range of each instrument. All the survey meters could obtain gamma-ray energy spectra, with an accuracy within ± 30% of the theoretical value. These survey meters are therefore applicable for radiation management.

元の言語英語
ページ(範囲)12-17
ページ数6
ジャーナルApplied Radiation and Isotopes
140
DOI
出版物ステータス出版済み - 10 1 2018

Fingerprint

scintillation
counters
pixels
photons
radiation
gamma ray spectra
linearity
energy spectra
dosage

All Science Journal Classification (ASJC) codes

  • Radiation

これを引用

Measurement of basic characteristics of scintillation-type radiation survey meters with multi-pixel photon counter. / Fujibuchi, Toshioh; Toyoda, Takatoshi; Terasaki, Kento.

:: Applied Radiation and Isotopes, 巻 140, 01.10.2018, p. 12-17.

研究成果: ジャーナルへの寄稿記事

@article{8e9eb47a1090416484ca59b57e44876a,
title = "Measurement of basic characteristics of scintillation-type radiation survey meters with multi-pixel photon counter",
abstract = "The basic characteristics of three low-cost radiation survey meters using multi-pixel photon counters (MPPC), the C12137, T-GMK2-S, and iMetry, were measured. The linearity of the dose rate was confirmed over the full range of each instrument. All the survey meters could obtain gamma-ray energy spectra, with an accuracy within ± 30{\%} of the theoretical value. These survey meters are therefore applicable for radiation management.",
author = "Toshioh Fujibuchi and Takatoshi Toyoda and Kento Terasaki",
year = "2018",
month = "10",
day = "1",
doi = "10.1016/j.apradiso.2018.06.010",
language = "English",
volume = "140",
pages = "12--17",
journal = "Applied Radiation and Isotopes",
issn = "0969-8043",
publisher = "Elsevier Limited",

}

TY - JOUR

T1 - Measurement of basic characteristics of scintillation-type radiation survey meters with multi-pixel photon counter

AU - Fujibuchi, Toshioh

AU - Toyoda, Takatoshi

AU - Terasaki, Kento

PY - 2018/10/1

Y1 - 2018/10/1

N2 - The basic characteristics of three low-cost radiation survey meters using multi-pixel photon counters (MPPC), the C12137, T-GMK2-S, and iMetry, were measured. The linearity of the dose rate was confirmed over the full range of each instrument. All the survey meters could obtain gamma-ray energy spectra, with an accuracy within ± 30% of the theoretical value. These survey meters are therefore applicable for radiation management.

AB - The basic characteristics of three low-cost radiation survey meters using multi-pixel photon counters (MPPC), the C12137, T-GMK2-S, and iMetry, were measured. The linearity of the dose rate was confirmed over the full range of each instrument. All the survey meters could obtain gamma-ray energy spectra, with an accuracy within ± 30% of the theoretical value. These survey meters are therefore applicable for radiation management.

UR - http://www.scopus.com/inward/record.url?scp=85048764757&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85048764757&partnerID=8YFLogxK

U2 - 10.1016/j.apradiso.2018.06.010

DO - 10.1016/j.apradiso.2018.06.010

M3 - Article

VL - 140

SP - 12

EP - 17

JO - Applied Radiation and Isotopes

JF - Applied Radiation and Isotopes

SN - 0969-8043