抄録
Radial profiles of density ratio of helium to hydrogen ions are measured using the charge exchange spectroscopy technique with the two-wavelength spectrometer system in the large helical device. The two-wavelength spectrometer system consists of a dichroic mirror box, a spectrometer with two grating and two camera lenses, and one CCD detector. The dichroic mirror box is used to divide the light of one fiber from the plasma to two fibers, one for HeII (λ = 468.6 nm) and the other for Hα (λ = 656.3 nm), that are connected to the entrance slit of the spectrometer to eliminate the interference between the HeII and the Hα spectra on the CCD. This system provides a simultaneous measurement of helium and hydrogen ion density ratio at 8 exact same locations (8 spatial channels) with a time resolution of >40 ms in the wide range of the density ratio of 0.05-5.
本文言語 | 英語 |
---|---|
論文番号 | 123514 |
ジャーナル | Review of Scientific Instruments |
巻 | 86 |
号 | 12 |
DOI | |
出版ステータス | 出版済み - 12月 2015 |
外部発表 | はい |
!!!All Science Journal Classification (ASJC) codes
- 器械工学