Measurement of response functions of a liquid organic scintillator for neutrons up to 800 MeV

Daiki Satoh, Tatsuhiko Sato, Akira Endo, Yasuhiro Yamaguchi, Masashi Takada, Kenji Ishibashi

研究成果: ジャーナルへの寄稿記事

25 引用 (Scopus)

抜粋

Response functions of a BC501A liquid organic scintillator for neutrons up to 800 MeV have been measured at the heavy-ion accelerator of the National Institute of Radiological Sciences, Japan. A thick graphite target was bombarded with 400-MeV/u C ions and 800-MeV/u Si ions to produce high-energy neutrons whose kinetic energy was determined by the time-of-flight method. The measured response functions were compared with the results obtained using SCINFUL-QMD code, and the accuracy of the code was experimentally verified up to 800 MeV. This work will contribute to extending the energies measurable with our new radiation dose-monitoring system (DARWIN), which is based on the BC501A scintillator.

元の言語英語
ページ(範囲)714-719
ページ数6
ジャーナルjournal of nuclear science and technology
43
発行部数7
DOI
出版物ステータス出版済み - 8 7 2006

All Science Journal Classification (ASJC) codes

  • Nuclear and High Energy Physics
  • Nuclear Energy and Engineering

フィンガープリント Measurement of response functions of a liquid organic scintillator for neutrons up to 800 MeV' の研究トピックを掘り下げます。これらはともに一意のフィンガープリントを構成します。

  • これを引用