Most instruments for measuring radon concentrations in the air should be calibrated using air with known radon concentrations, obtained from a solution of radium. However, safe handling of such a solution of alpha-active elements can be troublesome. Standard gamma-ray sources of low activity are handled more safely and typically used to obtain the absolute detection efficiency of a Ge detector with high accuracy. A new method has been developed to calibrate a radon detector with such sources. A radon exhalation rate for a substance containing a small amount of radium is measured with the radon detector to be calibrated. After this, the vessel containing the radium is sealed so that the radon does not escape from it. The buildup of the activity of 2 14Bi in it is obtained from gamma-ray measurements and gives the radon exhalation rate on the basis of the activity of the standard sources. From the comparison between the two values of the exhalation rate, the radon detector is calibrated. A plane multiwire-electrode ionization chamber is used as a radon detector, and its detection efficiency is calculated from its geometrical form. The radon exhalation rate computed from the calculated efficiency agrees with that determined from the activity of the standard gamma-ray sources within their specified 6% error.

本文言語英語
ページ(範囲)1189-1193
ページ数5
ジャーナルReview of Scientific Instruments
60
6
DOI
出版ステータス出版済み - 1989

All Science Journal Classification (ASJC) codes

  • 器械工学

フィンガープリント

「Method to calibrate an ionization chamber for measuring indoor radon concentrations with standard gamma-ray sources」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル