Online preconcentration by transient isotachophoresis in linear polymer on a poly(methyl methacrylate) microchip for separation of human serum albumin immunoassay mixtures

Mohamad Reza Mohamadi, Noritada Kaji, Manabu Tokeshi, Yoshinobu Baba

研究成果: ジャーナルへの寄稿学術誌査読

79 被引用数 (Scopus)

抄録

Online preconcentration of human serum albumin (HSA) and its immunocomplex with a monoclonal antibody by on-chip transient isotachophoresis is reported. An 800-fold signal enhancement was achieved following the preconcentration on standard cross-channel microchips made of poly (methyl methacrylate). Sample injection, preconcentration, and separation were done continuously and controlled solely by a sequential voltage switching program. The preconcentration was followed by on-chip nondenaturing gel electrophoresis in methylcellulose solution. The method was applied to microchip electrophoresis immunoassay of HSA. Baseline separation of HSA and its immunocomplex was achieved in 25 s in the first 1 cm of the microchannel. In a direct immunoassay, the minimum detectable concentration of fluorescent labeled HSA by laser-induced fluorescence detection was 7.5 pM.

本文言語英語
ページ(範囲)3667-3672
ページ数6
ジャーナルAnalytical Chemistry
79
10
DOI
出版ステータス出版済み - 5月 15 2007
外部発表はい

!!!All Science Journal Classification (ASJC) codes

  • 分析化学

フィンガープリント

「Online preconcentration by transient isotachophoresis in linear polymer on a poly(methyl methacrylate) microchip for separation of human serum albumin immunoassay mixtures」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル